The effect of various media and probe angles on the power output of the Cyclo G6 Glaucoma Laser System.
Patel, Shyam; Kakouri, Agni; Chaudhary, Shweta; Knepper, Paul A; Pfahler, Nicholas M; Samples, John; Giovingo, Michael.
Afiliação
  • Patel S; Department of Ophthalmology, John H. Stroger, Jr. Hospital of Cook County, Chicago, IL, USA.
  • Kakouri A; Department of Ophthalmology, John H. Stroger, Jr. Hospital of Cook County, Chicago, IL, USA.
  • Chaudhary S; Department of Ophthalmology and Visual Sciences, University of Illinois at Chicago, Chicago, IL, USA.
  • Knepper PA; Department of Ophthalmology, John H. Stroger, Jr. Hospital of Cook County, Chicago, IL, USA.
  • Pfahler NM; Department of Ophthalmology and Visual Sciences, University of Illinois at Chicago, Chicago, IL, USA.
  • Samples J; Feinberg School of Medicine, Department of Ophthalmology, Northwestern Medicine, Chicago, IL, USA.
  • Giovingo M; Department of Ophthalmology and Visual Sciences, University of Illinois at Chicago, Chicago, IL, USA.
Lasers Med Sci ; 36(3): 605-609, 2021 Apr.
Article em En | MEDLINE | ID: mdl-32681220
ABSTRACT
To evaluate the effect of various media and Iridex MicroPulse P3 (MP3) probe angles on the power output from the Cyclo G6 Glaucoma Laser (G6) System. A laser power meter was used to measure the power output (milliwatts, mW) of the Cyclo G6 System. Each of the ten trials consisted of measurements in six different media no substrate, balanced salt solution (BSS), artificial tears (AT), tetracaine eye drop, lubricating ointment, and lidocaine gel. The output of the MP3 probe was measured at an angle of 90° and 45°, submerged in the respective media. The output was also measured with the probe held at 90° but above the medium. The mean power outputs with the probe being held at 90° to the sensor with no substrate, BSS, AT, tetracaine eye drop, lubricating ointment, and lidocaine gel were 358 ± 16.8 mW, 612 ± 14.2 mW, 613 ± 13.3 mW, 612 ± 14.0 mW, 620 ± 9.9 mW, and 610 ± 12.2 mW, respectively. These values were statistically higher than noncontact and 45° probe angles for each medium. The values between any two media (excluding no substrate) at a 90° probe angle with full contact were not statistically significant. The highest output of the G6 System was obtained with a 90° probe angle, with full contact and any of the coupling media.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Glaucoma / Fotocoagulação a Laser / Eletricidade Idioma: En Ano de publicação: 2021 Tipo de documento: Article