Your browser doesn't support javascript.
loading
Long-term follow-up of the two-incision implantation technique for the subcutaneous implantable cardioverter-defibrillator.
van der Stuijt, Willeke; Baalman, Sarah W E; Brouwer, Tom F; Quast, Anne-Floor B E; de Groot, Joris R; Knops, Reinoud E.
Afiliação
  • van der Stuijt W; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.
  • Baalman SWE; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.
  • Brouwer TF; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.
  • Quast ABE; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.
  • de Groot JR; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.
  • Knops RE; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.
Pacing Clin Electrophysiol ; 43(12): 1476-1480, 2020 12.
Article em En | MEDLINE | ID: mdl-32720398
ABSTRACT

INTRODUCTION:

The two-incision implantation technique of the subcutaneous implantable cardioverter-defibrillator (S-ICD) was introduced as an alternative to the standard three-incision approach by omitting the superior parasternal incision. Thereby, complications may be prevented. Short-term follow-up demonstrated the safety and efficacy of the two-incision technique. However, long-term results are lacking.

METHODS:

This retrospective study included patients implanted between February 2009 and June 2020. Patients were divided into a group of patients who were implanted with the standard three-incision technique and a group who were implanted with the two-incision technique. Outcomes were defibrillation impedance and efficacy and complications requiring intervention.

RESULTS:

A total of 268 patients were included (age 42.4 ± 16.6 years, 35.4% female, BMI 25.1 ± 4.5 kg/m2 ). Thirty-one patients underwent S-ICD implantation with the three-incision technique and 237 patients with the two-incision technique. First shock efficacy during defibrillation testing was 93% in the three-incision group versus 94% in the two-incision group (P = .69), and shock impedance was 85 versus 68 ohms (P = .04). First shock success was 75% versus 76% for spontaneous episodes (P = 1.00). Complication-free survival at 5-year follow-up in the three-incision group was estimated at 0.96 (95% CI 0.90-1.00) versus 0.98 (95% CI 0.96-1.00) in the two-incision group (P = .20) and for inappropriate shocks at 5-year 0.77 (95% CI 0.63-0.94) versus 0.83 (95% CI 0.77-0.89, P = .30), respectively.

CONCLUSION:

Five-year follow-up in this S-ICD cohort showed similar complication rates and effectiveness of two-incision technique compared to the three-incision technique. This technique offers physicians a less invasive and more simplified implantation procedure for the S-ICD, with a better cosmetic result.