D-Dimer Varies Widely Across Instrument Platforms and is Not a Reliable Indicator of Periprosthetic Joint Infections.
Pearson, Lauren N; Moser, Karen A; Schmidt, Robert L.
Afiliação
  • Pearson LN; Department of Pathology, ARUP Laboratories & University of Utah, Salt Lake City, UT, USA.
  • Moser KA; Department of Pathology, ARUP Laboratories & University of Utah, Salt Lake City, UT, USA.
  • Schmidt RL; Department of Pathology, ARUP Laboratories & University of Utah, Salt Lake City, UT, USA.
Arthroplast Today ; 6(4): 686-688, 2020 Dec.
Article em En | MEDLINE | ID: mdl-32875020
ABSTRACT
The D-dimer test is a component of the modified scoring criteria for periprosthetic joint infection (PJI). The performance of the D-dimer test varies greatly among laboratories because of the lack of standardization. Laboratories may use different assays and will produce widely varying results for the same sample. This study used published proficiency testing data from 3903 laboratories to demonstrate the variability in D-dimer results and estimate the misclassification rate of patients using the proposed cutoff for the test as a component of PJI criteria. Given the variability in D-dimer results, a clinically significant percentage of patients are likely to be misclassified. The data illustrate that a universal cutoff for this marker in the context of assessment for PJI is not appropriate. Each site must conduct a study to determine an appropriate cutoff for their unique testing platform.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2020 Tipo de documento: Article