Your browser doesn't support javascript.
loading
An overview of the medical-physics-related verification system for radiotherapy multicenter clinical trials by the Medical Physics Working Group in the Japan Clinical Oncology Group-Radiation Therapy Study Group.
Nishio, Teiji; Nakamura, Mitsuhiro; Okamoto, Hiroyuki; Kito, Satoshi; Minemura, Toshiyuki; Ozawa, Shuichi; Kumazaki, Yu; Ishikawa, Masayori; Tohyama, Naoki; Kurooka, Masahiko; Nakashima, Takeo; Shimizu, Hidetoshi; Suzuki, Ryusuke; Ishikura, Satoshi; Nishimura, Yasumasa.
Afiliação
  • Nishio T; Department of Medical Physics, Graduate School of Medicine, Tokyo Women's Medical University, 8-1, Kawada-cho, Shinjuku-ku, Tokyo 162-8666, Japan.
  • Nakamura M; Medical Physics Working Group (MPWG) in Japan Clinical Oncology Group - Radiation Therapy Study Group (JCOG-RTSG), Tokyo, Japan.
  • Okamoto H; Division of Medical Physics, Department of Information Technology and Medical Engineering, Human He Sciences, Graduate School of Medicine, Kyoto University, 53 Shogoin-Kawaharacho, Sakyo-ku, Kyoto 606-8507, Japan.
  • Kito S; Medical Physics Working Group (MPWG) in Japan Clinical Oncology Group - Radiation Therapy Study Group (JCOG-RTSG), Tokyo, Japan.
  • Minemura T; Department of Medical Physics, National Cancer Center Hospital, 5-1-1 Tsukiji, Chuo-ku, Tokyo, 104-0045 Japan.
  • Ozawa S; Medical Physics Working Group (MPWG) in Japan Clinical Oncology Group - Radiation Therapy Study Group (JCOG-RTSG), Tokyo, Japan.
  • Kumazaki Y; Department of Radiology, Tokyo Metropolitan Bokutoh Hospital, 4-23-15 Kotobashi, Sumida-ku, Tokyo 130-8575, Japan.
  • Ishikawa M; Department of Radiation Oncology, Tokyo Metropolitan Cancer and Infectious Disease Center Komagome Hospital, 3-18-22 Honkomagome, Bunkyo-ku, Tokyo 113-8677, Japan.
  • Tohyama N; Division of Medical Physics, Department of Information Technology and Medical Engineering, Human He Sciences, Graduate School of Medicine, Kyoto University, 53 Shogoin-Kawaharacho, Sakyo-ku, Kyoto 606-8507, Japan.
  • Kurooka M; Medical Physics Working Group (MPWG) in Japan Clinical Oncology Group - Radiation Therapy Study Group (JCOG-RTSG), Tokyo, Japan.
  • Nakashima T; Division of Medical Support and Partnership, Center for Cancer Control and Information Services, National Cancer Center, 5-1-1 Tsukiji, Chuo-ku, Tokyo 104-0045, Japan.
  • Shimizu H; Medical Physics Working Group (MPWG) in Japan Clinical Oncology Group - Radiation Therapy Study Group (JCOG-RTSG), Tokyo, Japan.
  • Suzuki R; Department of Radiation Oncology, Graduate School of Biomedical and Health Sciences, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima 734-8551, Japan.
  • Ishikura S; Hiroshima High-Precision Radiotherapy Cancer Center, 3-2-2, Futabanosato, Higashi-ku, Hiroshima 732-0057, Japan.
  • Nishimura Y; Medical Physics Working Group (MPWG) in Japan Clinical Oncology Group - Radiation Therapy Study Group (JCOG-RTSG), Tokyo, Japan.
J Radiat Res ; 61(6): 999-1008, 2020 Nov 16.
Article em En | MEDLINE | ID: mdl-32989445
ABSTRACT
The Japan Clinical Oncology Group-Radiation Therapy Study Group (JCOG-RTSG) has initiated several multicenter clinical trials for high-precision radiotherapy, which are presently ongoing. When conducting multi-center clinical trials, a large difference in physical quantities, such as the absolute doses to the target and the organ at risk, as well as the irradiation localization accuracy, affects the treatment outcome. Therefore, the differences in the various physical quantities used in different institutions must be within an acceptable range for conducting multicenter clinical trials, and this must be verified with medical physics consideration. In 2011, Japan's first Medical Physics Working Group (MPWG) in the JCOG-RTSG was established to perform this medical-physics-related verification for multicenter clinical trials. We have developed an auditing method to verify the accuracy of the absolute dose and the irradiation localization. Subsequently, we credentialed the participating institutions in the JCOG multicenter clinical trials that were using stereotactic body radiotherapy (SBRT) for lungs, intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT) for several disease sites, and proton beam therapy (PT) for the liver. From the verification results, accuracies of the absolute dose and the irradiation localization among the participating institutions of the multicenter clinical trial were assured, and the JCOG clinical trials could be initiated.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Planejamento da Radioterapia Assistida por Computador / Ensaios Clínicos como Assunto / Radiocirurgia / Radioterapia de Intensidade Modulada Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Planejamento da Radioterapia Assistida por Computador / Ensaios Clínicos como Assunto / Radiocirurgia / Radioterapia de Intensidade Modulada Idioma: En Ano de publicação: 2020 Tipo de documento: Article