Mechanical Evaluation of the Stability of One or Two Miniscrews under Loading on Synthetic Bone.
Pradal, Andrea; Nucci, Ludovica; Derton, Nicola; De Felice, Maria Elena; Turco, Gianluca; Grassia, Vincenzo; Contardo, Luca.
Afiliação
  • Pradal A; Department of Medical, Surgical and Health Sciences, University of Trieste, 34127 Trieste, Italy.
  • Nucci L; Multidisciplinary Department of Medical-Surgical and Dental Specialties, University of Campania Luigi Vanvitelli, 81100 Naples, Italy.
  • Derton N; Department of Medical, Surgical and Health Sciences, University of Trieste, 34127 Trieste, Italy.
  • De Felice ME; Multidisciplinary Department of Medical-Surgical and Dental Specialties, University of Campania Luigi Vanvitelli, 81100 Naples, Italy.
  • Turco G; Department of Medical, Surgical and Health Sciences, University of Trieste, 34127 Trieste, Italy.
  • Grassia V; Multidisciplinary Department of Medical-Surgical and Dental Specialties, University of Campania Luigi Vanvitelli, 81100 Naples, Italy.
  • Contardo L; Department of Medical, Surgical and Health Sciences, University of Trieste, 34127 Trieste, Italy.
J Funct Biomater ; 11(4)2020 Nov 05.
Article em En | MEDLINE | ID: mdl-33167442
The aim of the present study was to evaluate the primary stability of a two-miniscrew system inserted into a synthetic bone and to compare the system with the traditional one. Forty-five bi-layered polyurethane blocks were used to simulate maxillary cancellous and cortical bone densities. Samples were randomly assigned to three groups-one-miniscrew system (Group A, N = 23), two-miniscrew system (Group B, N = 22) and archwire-only (Group C, N = 10). A total of 67 new miniscrews were subdivided into Group A (23 singles) and Group B (22 couples). 30 mm of 19″ × 25″ archwires were tied to the miniscrew. The load was applied perpendicularly to the archwire. Maximum Load Value (MLV), Yield Load (YL) and Loosening Load (LL) were recorded for each group. The YL of Group B and C had a mean value respectively of 4.189 ± 0.390 N and 3.652 ± 0.064 N. The MLV of Group A, B and C had a mean value respectively of 1.871 ± 0.318N, of 4.843 ± 0.515 N and 4.150 ± 0.086 N. The LL of Group A and B had a mean value respectively of 1.871 ± 0.318 N and of 2.294 ± 0.333 N. A two- temporary anchorage device (TAD) system is on average stiffer than a one-TAD system under orthodontic loading.