Your browser doesn't support javascript.
loading
The fatal consequence of inappropriate therapy in a single VF zone primary prevention defibrillator.
Griffiths, Samuel; Behar, Jonathan M; Lascelles, Karen; Wong, Tom.
Afiliação
  • Griffiths S; Department of Cardiology, Royal Brompton and Harefield NHS Foundation Trust, London, UK.
  • Behar JM; Department of Cardiology, Royal Brompton and Harefield NHS Foundation Trust, London, UK.
  • Lascelles K; Department of Cardiology, Royal Brompton and Harefield NHS Foundation Trust, London, UK.
  • Wong T; Department of Cardiology, Royal Brompton and Harefield NHS Foundation Trust, London, UK.
Pacing Clin Electrophysiol ; 44(4): 740-743, 2021 04.
Article em En | MEDLINE | ID: mdl-33386765
We report a case of inappropriate implantable cardioverter defibrillator (ICD) therapy due to single ventricular fibrillation (VF) zone programming leading to patient death. A remote transmission was received from a patient with a cardiac resynchronization therapy-defibrillator showing sinus tachycardia in the VF detection zone initiating inappropriate shocks and resulting in shock refractory VF. This case report highlights the importance of manufacturer specific ICD programming. In devices without discrimination in the VF zone, a higher rate single VF detection zone and/or addition of a ventricular tachycardia zone with supraventricular tachycardia discrimination should be considered.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fibrilação Ventricular / Dispositivos de Terapia de Ressincronização Cardíaca Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fibrilação Ventricular / Dispositivos de Terapia de Ressincronização Cardíaca Idioma: En Ano de publicação: 2021 Tipo de documento: Article