Your browser doesn't support javascript.
loading
Ergonomic effects of medical augmented reality glasses in video-assisted surgery.
Lim, Ae Kyeong; Ryu, Junsun; Yoon, Hong Man; Yang, Hee Chul; Kim, Seok-Ki.
Afiliação
  • Lim AK; Division of Convergence Technology, Research Institute of National Cancer Center, 323 Ilsan-ro, Ilsandong-gu, Goyang-si, Gyeonggi-do, 10408, Republic of Korea.
  • Ryu J; Division of Convergence Technology, Research Institute of National Cancer Center, 323 Ilsan-ro, Ilsandong-gu, Goyang-si, Gyeonggi-do, 10408, Republic of Korea.
  • Yoon HM; Center for Thyroid Cancer, 323, Ilsan-ro, Ilsandong-gu, Goyang-si, Gyeonggi-do, Republic of Korea.
  • Yang HC; Division of Convergence Technology, Research Institute of National Cancer Center, 323 Ilsan-ro, Ilsandong-gu, Goyang-si, Gyeonggi-do, 10408, Republic of Korea.
  • Kim SK; Center for Gastric Cancer, 323, Ilsan-ro, Ilsandong-gu, Goyang-si, Gyeonggi-do, Republic of Korea.
Surg Endosc ; 36(2): 988-998, 2022 02.
Article em En | MEDLINE | ID: mdl-33638103
ABSTRACT

BACKGROUND:

The aim of this study was to objectively compare medical augmented reality glasses (ARG) and conventional monitors in video-assisted surgery and to systematically analyze its ergonomic benefits.

METHODS:

Three surgeons (thoracic, laparoscopic, and thyroid surgeons) participated in the study. Six thoracoscopic metastasectomies, six subtotal laparoscopic gastrectomies, and six thyroidectomies were performed with and without ARG. The subjective experience was evaluated using a questionnaire-based NASA-Task Load Index (NASA-TLX). Postures during surgeries were recorded. The risk of musculoskeletal disorders associated with video-assisted surgery was assessed using rapid entire body assessment (REBA). Surface electromyography (EMG) was recorded. Muscle fatigue was objectively measured.

RESULTS:

NASA-TLX scores of three surgeons were lower when ARG was used compared to those with conventional monitor (66.4 versus 82.7). Less workload during surgery was reported with ARG. The laparoscopic surgeon exhibited a substantial decrease in mental and physical demand [- 21.1 and 12.5%)] and the thyroid surgeon did (- 40.0 and - 66.7%).Total REBA scores decreased with ARG (8 to 3.6). The risk of musculoskeletal disorders was improved in regions of the neck and shoulders. Root mean square (RMS) of the EMG signal decreased from 0.347 ± 0.150 to 0.286 ± 0.130 (p = 0.010) with usage of ARG; a decrease was observed in all surgeons. The greatest RMS decrease was observed in trapezius and sternocleidomastoid muscles. The decrease in brachioradialis muscle was small.

CONCLUSION:

ARG assisted with correction of bad posture in surgeons during video-assisted surgery and reduced muscular fatigue of the upper body. This study highlights the superior ergonomic efficiency of ARG in video-assisted surgery.