Performances of NeuMoDx™, a random-access system for hepatitis B virus DNA and hepatitis C virus RNA quantification.
Besombes, Juliette; Pronier, Charlotte; Lefevre, Charles; Lagathu, Gisèle; Maillard, Anne; Grolhier, Claire; Thibault, Vincent.
Afiliação
  • Besombes J; Virology Université Rennes, CHU Rennes, Inserm, EHESP, Irset (Institut de recherche en santé, environnement et travail)-UMR_S 1085, Rennes, France.
  • Pronier C; Virology Université Rennes, CHU Rennes, Inserm, EHESP, Irset (Institut de recherche en santé, environnement et travail)-UMR_S 1085, Rennes, France.
  • Lefevre C; Virology Université Rennes, CHU Rennes, Inserm, EHESP, Irset (Institut de recherche en santé, environnement et travail)-UMR_S 1085, Rennes, France.
  • Lagathu G; Virology Université Rennes, CHU Rennes, Inserm, EHESP, Irset (Institut de recherche en santé, environnement et travail)-UMR_S 1085, Rennes, France.
  • Maillard A; Virology Université Rennes, CHU Rennes, Inserm, EHESP, Irset (Institut de recherche en santé, environnement et travail)-UMR_S 1085, Rennes, France.
  • Grolhier C; Virology Université Rennes, CHU Rennes, Inserm, EHESP, Irset (Institut de recherche en santé, environnement et travail)-UMR_S 1085, Rennes, France.
  • Thibault V; Virology Université Rennes, CHU Rennes, Inserm, EHESP, Irset (Institut de recherche en santé, environnement et travail)-UMR_S 1085, Rennes, France. Electronic address: vincent.thibault@chu-rennes.fr.
Clin Microbiol Infect ; 27(11): 1693.e9-1693.e15, 2021 Nov.
Article em En | MEDLINE | ID: mdl-33677082
ABSTRACT

OBJECTIVE:

Monitoring of viral loads (VL) for hepatitis B and C viruses (HBV; HCV) is essential to evaluate disease progression and treatment response. Automated, random-access rapid systems are becoming standard to provide clinicians with reliable VL. The aim of this study was to evaluate the analytical performances of the recently launched NeuMoDx™ for HBV-DNA and HCV-RNA quantification.

METHODS:

Clinical samples routinely quantified on the Beckman-Veris system were either retrospectively (frozen samples; HBV n = 178, HCV n = 249), or in parallel (fresh primary tubes; HBV n = 103, HCV n = 117) tested using NeuMoDx™. Linearity range was assessed on serial dilutions of high-titre plasmas containing different genotypes for HBV (A-E, n = 9) and HCV (1a,1b,2-5, n = 12).

RESULTS:

Overall test failure, mostly internal control amplification failure, was 2.3% and was not influenced by matrix types (fresh or frozen). For HBV VL, κ agreement was 74%, with 27 (12.6%) discrepancies. Correlation between HBV assays on 72 quantified samples by both methods was excellent (r = 0.963) with a mean bias (NeuMoDx™-Veris) of 0.21 log IU/mL. For HCV VL, κ agreement reached 94%, with 9 (2.8%) discrepancies. The r correlation factor between assays on 104 samples was 0.960 with a mean bias of -0.14 log IU/mL (NeuMoDx™-Veris). Serial dilutions confirmed the claimed linear ranges for all analysed HBV and HCV genotypes. The mean turnaround time was 72 minutes (range 55-101 minutes) for HBV and 96 minutes (range 78-133 minutes) for HCV.

CONCLUSION:

Results obtained on the NeuMoDx™ confirmed the overall good functionality of the system with a short turn-around-time, full traceability and easy handling. These results on HBV and HCV VL look promising and should be challenged with further comparisons.