Intra- and inter-observer agreements in detecting peri-implant bone defects between periapical radiography and cone beam computed tomography: A clinical study.
Zhang, Chu-Nan; Zhu, Yu; Fan, Lin-Feng; Zhang, Xiao; Jiang, Yin-Hua; Gu, Ying-Xin.
Afiliação
  • Zhang CN; Department of Implant Dentistry, Shanghai Ninth People's Hospital, College of Stomatology, Shanghai Jiao Tong University School of Medicine; National Clinical Research Center for Oral Diseases; Shanghai Key Laboratory of Stomatology & Shanghai Research Institute of Stomatology, Shanghai, China.
  • Zhu Y; Department of Implant Dentistry, Shanghai Ninth People's Hospital, College of Stomatology, Shanghai Jiao Tong University School of Medicine; National Clinical Research Center for Oral Diseases; Shanghai Key Laboratory of Stomatology & Shanghai Research Institute of Stomatology, Shanghai, China.
  • Fan LF; Department of Radiology, Shanghai Ninth People's Hospital, College of Stomatology, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
  • Zhang X; Department of Implant Dentistry, Shanghai Ninth People's Hospital, College of Stomatology, Shanghai Jiao Tong University School of Medicine; National Clinical Research Center for Oral Diseases; Shanghai Key Laboratory of Stomatology & Shanghai Research Institute of Stomatology, Shanghai, China.
  • Jiang YH; Department of Oral and Maxillofacial Implantology, Lishui Sixth People's Hospital, Wenzhou Medical University, Lishui, China.
  • Gu YX; Department of Implant Dentistry, Shanghai Ninth People's Hospital, College of Stomatology, Shanghai Jiao Tong University School of Medicine; National Clinical Research Center for Oral Diseases; Shanghai Key Laboratory of Stomatology & Shanghai Research Institute of Stomatology, Shanghai, China.
J Dent Sci ; 16(3): 948-956, 2021 Jul.
Article em En | MEDLINE | ID: mdl-34141109
ABSTRACT
BACKGROUND/

PURPOSE:

Information regarding agreements between periapical radiograph (PA) and cone beam computed tomography (CBCT) in detecting peri-implant defect is still scarce. The aim of this clinical study was to compare agreements between PA and CBCT in detecting peri-implant bone defect. MATERIALS AND

METHODS:

This retrospective clinical study enrolled 32 patients with both PA and CBCT filmed right after implant placement. Four modalities were used for film reading PA1 (original), PA2 (enhanced brightness/contrast), CBCT1 (selected axial and mesial-distal direction images) and CBCT2 (all data with software). 2 experienced and 2 inexperienced observers scored all films. Intra- and inter-observer agreements were estimated with Cohen's kappa coefficient. Categorized agreements were compared and differences among four modalities were calculated.

RESULTS:

Agreements of PA were better than CBCT when detecting peri-implant bone defects in inter-observer agreements (median kappa 0.471 vs. 0.192; p = 0.016). Moreover, agreements in experienced observers were better than inexperienced observers (median kappa 0.883 vs. 0.567; p < 0.001). There was significant difference among four modalities except for experienced observer 2 (p = 0.218).

CONCLUSION:

Agreements of PA are better than CBCT when detecting peri-implant bone defects, especially for inter-observer agreements. Experienced observers are more consistent in assessment than inexperienced ones.