Your browser doesn't support javascript.
loading
CT-based thermometry with virtual monoenergetic images by dual-energy of fat, muscle and bone using FBP, iterative and deep learning-based reconstruction.
Heinrich, Andreas; Schenkl, Sebastian; Buckreus, David; Güttler, Felix V; Teichgräber, Ulf K-M.
Afiliação
  • Heinrich A; Department of Radiology, Jena University Hospital - Friedrich Schiller University, Am Klinikum 1, 07747, Jena, Germany. andreas.heinrich@med.uni-jena.de.
  • Schenkl S; Institute of Forensic Medicine, Jena University Hospital - Friedrich Schiller University, Am Klinikum 1, 07747, Jena, Germany.
  • Buckreus D; Department of Radiology, Jena University Hospital - Friedrich Schiller University, Am Klinikum 1, 07747, Jena, Germany.
  • Güttler FV; Department of Radiology, Jena University Hospital - Friedrich Schiller University, Am Klinikum 1, 07747, Jena, Germany.
  • Teichgräber UK; Department of Radiology, Jena University Hospital - Friedrich Schiller University, Am Klinikum 1, 07747, Jena, Germany.
Eur Radiol ; 32(1): 424-431, 2022 Jan.
Article em En | MEDLINE | ID: mdl-34327575
ABSTRACT

OBJECTIVES:

The aim of this study was to evaluate the sensitivity of CT-based thermometry for clinical applications regarding a three-component tissue phantom of fat, muscle and bone. Virtual monoenergetic images (VMI) by dual-energy measurements and conventional polychromatic 120-kVp images with modern reconstruction algorithms adaptive statistical iterative reconstruction-Volume (ASIR-V) and deep learning image reconstruction (DLIR) were compared.

METHODS:

A temperature-regulating water circuit system was developed for the systematic evaluation of the correlation between temperature and Hounsfield units (HU). The measurements were performed on a Revolution CT with gemstone spectral imaging technology (GSI). Complementary measurements were performed without GSI (voltage 120 kVp, current 130-545 mA). The measured object was a tissue equivalent phantom in a temperature range of 18 to 50°C. The evaluation was carried out for VMI at 40 to 140 keV and polychromatic 120-kVp images.

RESULTS:

The regression analysis showed a significant inverse linear dependency between temperature and average HU regardless of ASIR-V and DLIR. VMI show a higher temperature sensitivity compared to polychromatic images. The temperature sensitivities were 1.25 HU/°C (120 kVp) and 1.35 HU/°C (VMI at 140 keV) for fat, 0.38 HU/°C (120 kVp) and 0.47 HU/°C (VMI at 40 keV) for muscle and 1.15 HU/°C (120 kVp) and 3.58 HU/°C (VMI at 50 keV) for bone.

CONCLUSIONS:

Dual-energy with VMI enables a higher temperature sensitivity for fat, muscle and bone. The reconstruction with ASIR-V and DLIR has no significant influence on CT-based thermometry, which opens up the potential of drastic dose reductions. KEY POINTS • Virtual monoenergetic images (VMI) enable a higher temperature sensitivity for fat (8%), muscle (24%) and bone (211%) compared to conventional polychromatic 120-kVp images. • With VMI, there are parameters, e.g. monoenergy and reconstruction kernel, to modulate the temperature sensitivity. In contrast, there are no parameters to influence the temperature sensitivity for conventional polychromatic 120-kVp images. • The application of adaptive statistical iterative reconstruction-Volume (ASIR-V) and deep learning-based image reconstruction (DLIR) has no effect on CT-based thermometry, opening up the potential of drastic dose reductions in clinical applications.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Termometria / Aprendizado Profundo Idioma: En Ano de publicação: 2022 Tipo de documento: Article