Your browser doesn't support javascript.
loading
Evaluation of a Prototype of a Novel Galactomannan Sandwich Assay Using the VIDAS® Technology for the Diagnosis of Invasive Aspergillosis.
Gallet, Salomé; Garnaud, Cécile; Dragonetti, Céline; Rivoiron, Sophie; Cognet, Odile; Guo, Yuping; Lesénéchal, Mylène; Maubon, Danièle; Cornet, Muriel.
Afiliação
  • Gallet S; Laboratoire de Parasitologie-Mycologie, Département des Agents Infectieux, Institut de Biologie et de Pathologie, CHU Grenoble Alpes, Grenoble, France.
  • Garnaud C; Laboratoire de Parasitologie-Mycologie, Département des Agents Infectieux, Institut de Biologie et de Pathologie, CHU Grenoble Alpes, Grenoble, France.
  • Dragonetti C; Univ. Grenoble Alpes, CNRS, CHU Grenoble Alpes, Grenoble INP, TIMC, Grenoble, France.
  • Rivoiron S; Department of Immunoassays, R&D bioMérieux, Marcy l'Etoile, France.
  • Cognet O; Department of Immunoassays, R&D bioMérieux, Marcy l'Etoile, France.
  • Guo Y; Laboratoire de Parasitologie-Mycologie, Département des Agents Infectieux, Institut de Biologie et de Pathologie, CHU Grenoble Alpes, Grenoble, France.
  • Lesénéchal M; Department of Immunoassays, R&D bioMérieux, Marcy l'Etoile, France.
  • Maubon D; Department of Immunoassays, R&D bioMérieux, Marcy l'Etoile, France.
  • Cornet M; Laboratoire de Parasitologie-Mycologie, Département des Agents Infectieux, Institut de Biologie et de Pathologie, CHU Grenoble Alpes, Grenoble, France.
Front Cell Infect Microbiol ; 11: 669237, 2021.
Article em En | MEDLINE | ID: mdl-34336710
ABSTRACT

Objectives:

To evaluate the analytical and clinical performance of a prototype of a VIDAS® Galactomannan (GM) unitary test (bioMérieux, Marcy l'Etoile, France) and compare to that of the Platelia™ Aspergillus Ag assay (Bio-Rad, CA, USA).

Methods:

Repeatability, reproducibility, and freeze-thaw stability of VIDAS®GM were evaluated. Sera from patients at risk of IA were concurrently tested with both the VIDAS®GM and Platelia™ Aspergillus Ag assays. Correlations between the two assays were assessed by Passing Bablok (PB) regression and performance by ROC analysis.

Results:

The correlations between the VIDAS®GM indexes after one and two cycles of freezing/thawing were r=1.00 and r=0.989, respectively. The coefficients of variation for negative, low-positive, and positive sera were 13%, 6%, and 5% for repeatability and 14.4%, 7.2%, and 5.5% for reproducibility. Overall, 126 sera were tested with both assays (44 fresh and 82 frozen). The correlation between VIDAS®GM and Platelia™ Aspergillus Ag was r=0.798. The areas under the curve of the ROC analyses were 0.892 and 0.894, for VIDAS®GM and Platelia™ Aspergillus Ag, respectively.

Conclusions:

This new VIDAS®GM prototype assay showed adequate analytical and clinical performance and a good correlation with that of Platelia™ Aspergillus Ag with 126 sera, although these results need to be confirmed in a larger prospective and multicentric study. As for the other VIDAS® assays, VIDAS®GM is a single-sample automated test using a solid reagent strip and receptacle. It is easy to use and suitable for rapid on-demand test results.