Diagnostic performance of an algorithm for automated large vessel occlusion detection on CT angiography.
Luijten, Sven P R; Wolff, Lennard; Duvekot, Martijne H C; van Doormaal, Pieter-Jan; Moudrous, Walid; Kerkhoff, Henk; Lycklama A Nijeholt, Geert J; Bokkers, Reinoud P H; Yo, Lonneke S F; Hofmeijer, Jeannette; van Zwam, Wim H; van Es, Adriaan C G M; Dippel, Diederik W J; Roozenbeek, Bob; van der Lugt, Aad.
Afiliação
  • Luijten SPR; Radiology & Nuclear Medicine, Erasmus MC University Medical Center, Rotterdam, The Netherlands s.luijten@erasmusmc.nl.
  • Wolff L; Radiology & Nuclear Medicine, Erasmus MC University Medical Center, Rotterdam, The Netherlands.
  • Duvekot MHC; Neurology, Erasmus MC University Medical Center, Rotterdam, The Netherlands.
  • van Doormaal PJ; Neurology, Albert Schweitzer Hospital, Dordrecht, The Netherlands.
  • Moudrous W; Radiology & Nuclear Medicine, Erasmus MC University Medical Center, Rotterdam, The Netherlands.
  • Kerkhoff H; Neurology, Maasstad Ziekenhuis, Rotterdam, Zuid-Holland, The Netherlands.
  • Lycklama A Nijeholt GJ; Neurology, Albert Schweitzer Hospital, Dordrecht, The Netherlands.
  • Bokkers RPH; Radiology, HMC Westeinde, Den Haag, Zuid-Holland, The Netherlands.
  • Yo LSF; Radiology, University Medical Center Groningen, Groningen, The Netherlands.
  • Hofmeijer J; Radiology, Catharina Hospital, Eindhoven, The Netherlands.
  • van Zwam WH; Neurology, Rijnstate Hospital, Arnhem, The Netherlands.
  • van Es ACGM; Radiology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Dippel DWJ; Radiology, Leiden University Medical Center, Leiden, Zuid-Holland, The Netherlands.
  • Roozenbeek B; Neurology, Erasmus MC University Medical Center, Rotterdam, The Netherlands.
  • van der Lugt A; Radiology & Nuclear Medicine, Erasmus MC University Medical Center, Rotterdam, The Netherlands.
J Neurointerv Surg ; 14(8): 794-798, 2022 Aug.
Article em En | MEDLINE | ID: mdl-34413245
BACKGROUND: Machine learning algorithms hold the potential to contribute to fast and accurate detection of large vessel occlusion (LVO) in patients with suspected acute ischemic stroke. We assessed the diagnostic performance of an automated LVO detection algorithm on CT angiography (CTA). METHODS: Data from the MR CLEAN Registry and PRESTO were used including patients with and without LVO. CTA data were analyzed by the algorithm for detection and localization of LVO (intracranial internal carotid artery (ICA)/ICA terminus (ICA-T), M1, or M2). Assessments done by expert neuroradiologists were used as reference. Diagnostic performance was assessed for detection of LVO and per occlusion location by means of sensitivity, specificity, and area under the curve (AUC). RESULTS: We analyzed CTAs of 1110 patients from the MR CLEAN Registry (median age (IQR) 71 years (60-80); 584 men; 1110 with LVO) and of 646 patients from PRESTO (median age (IQR) 73 years (62-82); 358 men; 141 with and 505 without LVO). For detection of LVO, the algorithm yielded a sensitivity of 89% in the MR CLEAN Registry and a sensitivity of 72%, specificity of 78%, and AUC of 0.75 in PRESTO. Sensitivity per occlusion location was 88% for ICA/ICA-T, 94% for M1, and 72% for M2 occlusion in the MR CLEAN Registry, and 80% for ICA/ICA-T, 95% for M1, and 49% for M2 occlusion in PRESTO. CONCLUSION: The algorithm provided a high detection rate for proximal LVO, but performance varied significantly by occlusion location. Detection of M2 occlusion needs further improvement.