Your browser doesn't support javascript.
loading
Neutral Nonenzymatic Glucose Biosensors Based on Electrochemically Deposited Pt/Au Nanoalloy Electrodes.
Lin, Fang-Yu; Lee, Pei-Yuan; Chu, Tien-Fu; Peng, Chang-I; Wang, Gou-Jen.
Afiliação
  • Lin FY; Graduate Institute of Biomedical Engineering, National Chung-Hsing University, Taichung, 40227, Taiwan.
  • Lee PY; Graduate Institute of Biomedical Engineering, National Chung-Hsing University, Taichung, 40227, Taiwan.
  • Chu TF; Department of Orthopedics, Show Chwan Memorial Hospital, Changhua, 50008, Taiwan.
  • Peng CI; Department of Mechanical Engineering, National Chung-Hsing University, Taichung, 40227, Taiwan.
  • Wang GJ; Department of Mechanical Engineering, National Chung-Hsing University, Taichung, 40227, Taiwan.
Int J Nanomedicine ; 16: 5551-5563, 2021.
Article em En | MEDLINE | ID: mdl-34429599
ABSTRACT

BACKGROUND:

Type I diabetes occurs when the pancreas can only make limited or minimal insulin. Patients with type 1 diabetes need effective approaches to manage diabetes and maintain their blood-glucose concentration. Recently, continuous glucose monitoring (CGM) has been used to help control blood-glucose levels in patients with type 1 diabetes. Patients with type 2 diabetes may also benefit from CGM on multiple insulin injections, basal insulin, or sulfonylureas. Enzyme-free glucose detection in a neutral environment is the recent development trend of CGM. MATERIALS AND

METHODS:

Pt/Au alloy electrodes for enzyme-free glucose detection in a neutral environment were formed by electrochemically depositing Pt/Au alloy on a thin polycarbonate (PC) membrane surface with a uniformly distributed micro-hemisphere array. The PC membranes were fabricated using semiconductor microelectromechanical manufacturing processes, precision micro-molding, and hot embossing. Amperometry was used to measure the glucose concentration in PBS (pH 7.4) and artificial human serum.

RESULTS:

The Pt/Au nanoalloy electrode had excellent specificity for glucose detection, according to the experimental results. The device had a sensitivity of 2.82 µA mM-1 cm-2, a linear range of 1.39-13.9 mM, and a detection limit of 0.482 mM. Even though the complex interfering species in human blood can degrade the sensing signal, further experiments conducted in artificial serum confirmed the feasibility of the proposed Pt/Au nanoalloy electrode in clinical applications.

CONCLUSION:

The proposed Pt/Au nanoalloy electrode can catalyze glucose reactions in neutral solutions with enhancing sensing performance by the synergistic effect of bimetallic materials and increasing detection surface area. This novel glucose biosensor has advantages, such as technology foresight, good detection performance, and high mass production feasibility. Thus, the proposed neutral nonenzymatic glucose sensor can be further used in CGMs.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Técnicas Biossensoriais / Diabetes Mellitus Tipo 2 Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Técnicas Biossensoriais / Diabetes Mellitus Tipo 2 Idioma: En Ano de publicação: 2021 Tipo de documento: Article