How biocompatible haemodialysers can conquer the need for systemic anticoagulation even in post-dilution haemodiafiltration: a cross-over study.
Vanommeslaeghe, Floris; Josipovic, Iván; Boone, Matthieu; van der Tol, Arjan; Dhondt, Annemie; Van Biesen, Wim; Eloot, Sunny.
Afiliação
  • Vanommeslaeghe F; Department of Nephrology, Ghent University Hospital, Ghent, Belgium.
  • Josipovic I; Centre for X-ray Tomography, Physics and Astronomy, Ghent University, Ghent, Belgium.
  • Boone M; Centre for X-ray Tomography, Physics and Astronomy, Ghent University, Ghent, Belgium.
  • van der Tol A; Department of Nephrology, Ghent University Hospital, Ghent, Belgium.
  • Dhondt A; Department of Nephrology, Ghent University Hospital, Ghent, Belgium.
  • Van Biesen W; Department of Nephrology, Ghent University Hospital, Ghent, Belgium.
  • Eloot S; Department of Nephrology, Ghent University Hospital, Ghent, Belgium.
Clin Kidney J ; 14(7): 1752-1759, 2021 Jul.
Article em En | MEDLINE | ID: mdl-34548919
ABSTRACT

BACKGROUND:

While systemic anticoagulation is most widely used in haemodialysis (HD), contraindications to its use might occur in particular settings. The Solacea™ haemodialyser with an asymmetric triacetate membrane claims improved biocompatibility and has already shown promising results when used in combination with only half dose of anticoagulation. To quantify the performance of the Solacea™ when further decreasing anticoagulation to zero, fibre blocking was assessed by micro-computed tomography (micro-CT).

METHODS:

Ten maintenance HD patients underwent six dialysis sessions at midweek using a Solacea™ 19H dialyser, consecutively in pre-dilution haemodiafiltration (pre-HDF), HD and post-dilution HDF (post-HDF). After the first three sessions with only a quarter of their regular anticoagulation dose (one-quarter), the last three sessions were performed without anticoagulation (zero). Dialyser fibre blocking was quantified in the dialyser outlet potting using a 3D micro-CT scanning technique post-dialysis.

RESULTS:

Even in case of reduced (one-quarter) anticoagulation, the relative number of open fibres post-dialysis was almost optimal, i.e. 0.96 (0.87-0.99) with pre-HDF, 0.99 (0.97-0.99) with HD and 0.97 (0.92-0.99) with post-HDF. Fibre patency was mildly decreased for pre-HDF and HD when anticoagulation was decreased from one-quarter to zero, i.e. to 0.76 (0.61-0.85) with pre-HDF (P = 0.004) and to 0.80 (0.77-0.89) with HD (P = 0.013). Comparing the results for zero anticoagulation, post-HDF [i.e. 0.94 (0.82-0.97)] performed as well as HD and pre-HDF.

CONCLUSIONS:

The Solacea™ dialyser provides promising results for use in conditions where systemic anticoagulation is contraindicated. Post-HDF, although inducing haemoconcentration in the dialyser, is equally effective for fibre patency in case of zero anticoagulation as pre-HDF and HD when using Solacea™.