Your browser doesn't support javascript.
loading
Prediction Algorithms for Blood Pressure Based on Pulse Wave Velocity Using Health Checkup Data in Healthy Korean Men: Algorithm Development and Validation.
Park, Dohyun; Cho, Soo Jin; Kim, Kyunga; Woo, Hyunki; Kim, Jee Eun; Lee, Jin-Young; Koh, Janghyun; Lee, JeanHyoung; Choi, Jong Soo; Chang, Dong Kyung; Choi, Yoon-Ho; Chung, Ji In; Cha, Won Chul; Jeong, Ok Soon; Jekal, Se Yong; Kang, Mira.
Afiliação
  • Park D; Department of Digital Health, Samsung Advanced Institute of Health Sciences and Technology, Sungkyunkwan University, Seoul, Republic of Korea.
  • Cho SJ; Center for Health Promotion, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Kim K; Department of Digital Health, Samsung Advanced Institute of Health Sciences and Technology, Sungkyunkwan University, Seoul, Republic of Korea.
  • Woo H; Statistics and Data Center, Research Institute for Future Medicine, Samsung Medical Center, Seoul, Republic of Korea.
  • Kim JE; Data Science Team, Evidnet Inc, Gyeonggi-do, Republic of Korea.
  • Lee JY; Center for Health Promotion, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Koh J; Center for Health Promotion, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Lee J; Center for Health Promotion, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Choi JS; Digital Innovation Center, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Chang DK; Digital Innovation Center, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Choi YH; Department of Digital Health, Samsung Advanced Institute of Health Sciences and Technology, Sungkyunkwan University, Seoul, Republic of Korea.
  • Chung JI; Division of Gastroenterology, Department of Internal Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Cha WC; Center for Health Promotion, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Jeong OS; Center for Health Promotion, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Jekal SY; Department of Digital Health, Samsung Advanced Institute of Health Sciences and Technology, Sungkyunkwan University, Seoul, Republic of Korea.
  • Kang M; Digital Innovation Center, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
JMIR Med Inform ; 9(12): e29212, 2021 Dec 08.
Article em En | MEDLINE | ID: mdl-34889753
BACKGROUND: Pulse transit time and pulse wave velocity (PWV) are related to blood pressure (BP), and there were continuous attempts to use these to predict BP through wearable devices. However, previous studies were conducted on a small scale and could not confirm the relative importance of each variable in predicting BP. OBJECTIVE: This study aims to predict systolic blood pressure and diastolic blood pressure based on PWV and to evaluate the relative importance of each clinical variable used in BP prediction models. METHODS: This study was conducted on 1362 healthy men older than 18 years who visited the Samsung Medical Center. The systolic blood pressure and diastolic blood pressure were estimated using the multiple linear regression method. Models were divided into two groups based on age: younger than 60 years and 60 years or older; 200 seeds were repeated in consideration of partition bias. Mean of error, absolute error, and root mean square error were used as performance metrics. RESULTS: The model divided into two age groups (younger than 60 years and 60 years and older) performed better than the model without division. The performance difference between the model using only three variables (PWV, BMI, age) and the model using 17 variables was not significant. Our final model using PWV, BMI, and age met the criteria presented by the American Association for the Advancement of Medical Instrumentation. The prediction errors were within the range of about 9 to 12 mmHg that can occur with a gold standard mercury sphygmomanometer. CONCLUSIONS: Dividing age based on the age of 60 years showed better BP prediction performance, and it could show good performance even if only PWV, BMI, and age variables were included. Our final model with the minimal number of variables (PWB, BMI, age) would be efficient and feasible for predicting BP.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2021 Tipo de documento: Article