Automated seizure detection in an EMU setting: Are software packages ready for implementation?
Reus, E E M; Visser, G H; van Dijk, J G; Cox, F M E.
Afiliação
  • Reus EEM; Department of Clinical Neurophysiology, Stichting Epilepsie Instellingen Nederland, the Netherlands. Electronic address: ereus@sein.nl.
  • Visser GH; Department of Clinical Neurophysiology, Stichting Epilepsie Instellingen Nederland, the Netherlands.
  • van Dijk JG; Department of Neurology, Leiden University Medical Centre, Leiden, the Netherlands.
  • Cox FME; Department of Clinical Neurophysiology, Stichting Epilepsie Instellingen Nederland, the Netherlands.
Seizure ; 96: 13-17, 2022 Mar.
Article em En | MEDLINE | ID: mdl-35042003
ABSTRACT

PURPOSE:

We assessed whether automated detection software, combined with live observation, enabled reliable seizure detection using three commercial software packages Persyst, Encevis and BESA.

METHODS:

Two hundred and eighty-six prolonged EEG records of individuals aged 16-86 years, collected between August 2019 and January 2020, were retrospectively processed using all three packages. The reference standard included all seizures mentioned in the clinical report supplemented with true detections made by the software and not previously detected by clinical physiologists. Sensitivity was measured for offline review by clinical physiologists and software seizure detection, both in combination with live monitoring in an EMU setting, for all three software packages at record and seizure level.

RESULTS:

The database contained 249 seizures in 64 records. The sensitivity of seizure detection was 98% for Encevis and Persyst, and 95% for BESA, when a positive results was defined as detection at least one of the seizures occurring within an individual record. When positivity was defined as recognition of all seizures, sensitivity was 93% for Persyst, 88% for Encevis and 84% for BESA. Clinical physiologists' review had a sensitivity of 100% at record level and 98% at seizure level. The median false positive rate per record was 1.7 for Persyst, 2.4 for BESA and 5.5 for Encevis per 24 h.

CONCLUSION:

Automated seizure detection software does not perform as well as technicians do. However, it can be used in an EMU setting when the user is aware of its weaknesses. This assessment gives future users helpful insight into these strengths and weaknesses. The Persyst software performs best.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dromaiidae Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dromaiidae Idioma: En Ano de publicação: 2022 Tipo de documento: Article