Comparison of the Automated Pattern-Noise (PANO) Glaucoma Test with the HFA, an FDT Stimulus, and the Fundus Area Cup-to-disk Ratio.
Hannen, Thomas; El-Khoury, Sylvain; Patel, Rajesh; Ngounou, Faustin; Preußner, Paul-Rolf.
Afiliação
  • Hannen T; Department of Ophthalmology, University Medical Center Mainz, Langenbeckstraße 1, Mainz, Germany.
  • El-Khoury S; Department of Ophthalmology, University Medical Center Mainz, Langenbeckstraße 1, Mainz, Germany; Department of Pediatrics and Retina, Fondation Ophtalmologique Adolphe de Rothschild, Paris, France.
  • Patel R; Department of Ophthalmology, Presbyterian Eye Hospital, Bafoussam, Cameroon.
  • Ngounou F; Department of Ophthalmology, Presbyterian Eye Hospital, Bafoussam, Cameroon.
  • Preußner PR; Department of Ophthalmology, University Medical Center Mainz, Langenbeckstraße 1, Mainz, Germany.
J Curr Glaucoma Pract ; 15(3): 132-138, 2021.
Article em En | MEDLINE | ID: mdl-35173395
ABSTRACT
AIM AND

OBJECTIVE:

To compare the results of a new automated glaucoma test-Pattern-Noise (PANO)-to the Humphrey Visual Field Analyzer-II (HFA), the fundus area cup-to-disk ratio (CDR), and a frequency doubling technology (FDT) stimulus. MATERIALS AND

METHODS:

This was a prospective study performed in the West-Region of Cameroon. Two hundred and nineteen eyes of 122 adult patients were included with a clinical suspicion of normal-tension or primary open-angle glaucoma and no other major ocular pathology. Eyes were examined with PANO, HFA (24-2 SITA standard), and FDT-stimulus in a randomized order followed by clinical assessment of the CDR.

RESULTS:

Parametric correlation of the mean contrast threshold of PANO with the mean contrast threshold of FDT-stimulus, total deviation of HFA, and area CDR was 0.94, -0.85, and 0.62, respectively (p < 0.001 for all values). Spatial distribution of sensitivity thresholds is highly correlated (p < 0.001) at all points in the visual field between PANO and HFA. With cut-off values of 3 ± 1 dB for HFA mean deviation and 4 ± 1 for PANO mean contrast threshold and after eliminating borderline cases, PANO's sensitivity was 95% and specificity 60%. The mean patient age was 45.2 ± 15.8 years. Mean thresholds of PANO and FDT-stimulus decreased with increasing age. Mean examination time was 7.1 ± 1.8 minutes for PANO, 5.9 ± 1.3 minutes for HFA, and 4.7 ± 1.3 minutes for FDT-stimulus. The mean percentage of false-positives per examination was 4.95% for PANO, 4.62% (p = 0.025) for FDT-stimulus, and 2.10% for HFA.

CONCLUSION:

The results showed that PANO was successful in suspecting the presence of glaucoma. Pattern-Noise examination led to findings that were significantly correlated to HFA, FDT stimulus, and area CDR. Some patterns of defect were also correlated. Furthermore, PANO showed a reasonable examination time and error rate. CLINICAL

SIGNIFICANCE:

Affordable and robust visual field devices are lacking in large parts of the developing world. Comparing them to established methods is a prerequisite to their clinical use. HOW TO CITE THIS ARTICLE Hannen T, El-Khoury S, Patel R, et al. Comparison of the Automated Pattern-Noise (PANO) Glaucoma Test with the HFA, an FDT Stimulus, and the Fundus Area Cup-to-disk Ratio. J Curr Glaucoma Pract 2021;15(3)132-138.