Dosimetric feasibility of hippocampal avoidance whole brain radiotherapy with an MRI-guided linear accelerator.
J Appl Clin Med Phys
; 23(6): e13587, 2022 Jun.
Article
em En
| MEDLINE
| ID: mdl-35344266
ABSTRACT
PURPOSE/OBJECTIVE(S) Whole brain radiotherapy with hippocampal avoidance (HA-WBRT) is a technique utilized to treat metastatic brain disease while preserving memory and neurocognitive function. We hypothesized that the treatment planning and delivery of HA-WBRT plans is feasible with an MRI-guided linear accelerator (linac) and compared plan results with clinical non-MRI-guided C-Arm linac plans. MATERIALS/METHODS:
Twelve HA-WBRT patients treated on a non-MRI-guided C-Arm linac were selected for retrospective analysis. Treatment plans were developed using a 0.35T MRI-guided linac system for comparison to clinical plans. Treatment planning goals were defined as provided in the Phase II Trial NRG CC001. MRI-guided radiotherapy (MRgRT) treatment plans were developed by a dosimetrist and compared with clinical plans. quality assurance (QA) plans were generated and delivered on the MRI-guided linac to a cylindrical diode detector array. Planning target volume (PTV) coverage was normalized to â¼95% to provide a control point for comparison of dose to the organs at risk.RESULTS:
MRgRT plans were deliverable and met all clinical goals. Mean values demonstrated that the clinical plans were less heterogeneous than MRgRT plans with mean PTV V37.5 Gy of 0.00% and 0.03% (p = 0.013), respectively. Average hippocampi maximum doses were 14.19 ± 1.29 Gy and 15.00 ± 1.51 Gy, respectively. The gamma analysis comparing planned and measured doses resulted in a mean of 99.9% ± 0.12% of passing points (3%/2mm criteria). MRgRT plans had an average of 38.33 beams with average total delivery time and beam-on time of 13.7 (11.2-17.5) min and 4.1 (3.2-5.4) min, respectively. Clinical plan delivery times ranged from 3 to 7 min depending on the number of noncoplanar arcs. Planning time between the clinical and MRgRT plans was comparable.CONCLUSION:
This study demonstrates that HA-WBRT can be treated using an MRI-guided linear accelerator with comparable treatment plan quality and delivery accuracy.Palavras-chave