Your browser doesn't support javascript.
loading
Structural failure in bridging stentgrafts for branched endovascular aneurysm repair: a case-control study.
Hauck, Sven R; Kupferthaler, Alexander; Freund, Martin C; Pichler, Peter; Stelzmüller, Marie-Elisabeth; Burghuber, Christopher; Ehrlich, Marek; Teufelsbauer, Harald; Loewe, Christian; Funovics, Martin A.
Afiliação
  • Hauck SR; Division of Cardiovascular and Interventional Radiology, Department of Bioimaging and Image-Guided Therapy, Medical University of Vienna, Vienna, Austria.
  • Kupferthaler A; Division of Cardiovascular and Interventional Radiology, Department of Bioimaging and Image-Guided Therapy, Medical University of Vienna, Vienna, Austria.
  • Freund MC; Department of Diagnostic and Interventional Radiology, Ordensklinikum Linz & Medical Faculty, Johannes Kepler University, Linz, Austria.
  • Pichler P; Department of Radiology, Medical University Innsbruck and Tirol Kliniken GmbH, Innsbruck, Austria.
  • Stelzmüller ME; Department of Radiology, Johannes Kepler University, Linz, Austria.
  • Burghuber C; Department of Cardiothoracic Surgery, Medical University of Vienna, Vienna, Austria.
  • Ehrlich M; Division of Vascular Surgery, Department of Surgery, Medical University of Vienna, Vienna, Austria.
  • Teufelsbauer H; Department of Cardiothoracic Surgery, Medical University of Vienna, Vienna, Austria.
  • Loewe C; Division of Vascular Surgery, Department of Surgery, Medical University of Vienna, Vienna, Austria.
  • Funovics MA; Division of Cardiovascular and Interventional Radiology, Department of Bioimaging and Image-Guided Therapy, Medical University of Vienna, Vienna, Austria.
Insights Imaging ; 13(1): 62, 2022 Mar 28.
Article em En | MEDLINE | ID: mdl-35347467
ABSTRACT

OBJECTIVE:

To present a case series of spontaneous structural failure of bridging stentgrafts (BSGs) after branched endovascular aortic repair (bEVAR), as well as their failure types and their detection. While bEVAR is a safe and effective procedure, one main limitation is the reintervention rate associated with the BSGs. Structural failure of BSGs, defined as fabric disruption, stent fracture with leak or complete separation is a major cause for reinterventions and difficult to detect in computed tomography angiography (CTA).

METHODS:

From a multicenter bEVAR complication database, structural BSG failures were identified. Patient and BSG characteristics, detection mode, failure type, treatment and outcome were recorded and compared with bEVAR patients with intact BSGs.

RESULTS:

Twenty-three BSG failures were detected in 12 patients with only 43% directly identified in CTA, after a mean of 21.4 months after implantation. The BSGs were Advanta (n = 4), E-Ventus (n = 16) and BeGraft (n = 3) in 10 renal, 9 superior mesenteric, and 4 celiac branches. Religning with another BSG was successful in 20/22 cases, one BSG failure necessitated renal branch embolization (organ loss), and one mesenteric bypass surgery. Two reintervention-related mortalities occurred.

CONCLUSION:

Structural failure of BSGs is a serious limitation for bEVAR, which can result in high reintervention rates and serious complications. BSG failure typically occurs in single-layer types and events are clustered in patients. The necessary reinterventions carry serious morbidity and mortality. Since the use as BSG in bEVAR is off-label with all current BSG manufacturers, caution is advised regarding patient-informed consent.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2022 Tipo de documento: Article