Your browser doesn't support javascript.
loading
Relationship Between Glottic View and Intubation Force During Macintosh and Airtraq Laryngoscopy and Intubation.
Hindman, Bradley J; Dexter, Franklin; Gadomski, Benjamin C; Puttlitz, Christian M.
Afiliação
  • Hindman BJ; From the Department of Anesthesia, University of Iowa Roy J. and Lucille A. Carver College of Medicine, Iowa City, Iowa.
  • Dexter F; From the Department of Anesthesia, University of Iowa Roy J. and Lucille A. Carver College of Medicine, Iowa City, Iowa.
  • Gadomski BC; Department of Mechanical Engineering, School of Biomedical Engineering, Orthopaedic Bioengineering Research Laboratory, Colorado State University, Fort Collins, Colorado.
  • Puttlitz CM; Department of Mechanical Engineering, School of Biomedical Engineering, Orthopaedic Bioengineering Research Laboratory, Colorado State University, Fort Collins, Colorado.
Anesth Analg ; 135(4): 815-819, 2022 10 01.
Article em En | MEDLINE | ID: mdl-35551148
ABSTRACT

BACKGROUND:

Because intubation-mediated cervical spine and spinal cord injury are likely determined by intubation force magnitude, understanding the determinants of intubation force magnitude is clinically relevant. With direct (Macintosh) laryngoscopy, when glottic view is less favorable, anesthesiologists apply greater force. We hypothesized that, when compared with direct (Macintosh) laryngoscopy, intubation force with an optical indirect laryngoscope (Airtraq) would be less dependent on glottic visualization.

METHODS:

Using data obtained in a prior clinical study, we tested whether the slope of the intubation force versus glottic view relationship differed between intubations performed in 14 patients who were intubated twice, once with a Macintosh and once with an Airtraq videolaryngoscope. Slopes were compared using least-squares linear regression and robust regression.

RESULTS:

The slope of the intubation force (N) versus glottic view (%) relationship with the Macintosh (-0.679 [standard error {SE}, 0.147]) was significantly more negative than that of the Airtraq (-0.076 [SE, 0.246]). The least-squares regression difference in slopes was -0.603 (SE, 0.287); P = .046. The robust regression difference in slopes was -0.747 (SE, 0.187); P = .0005. Thus, when compared with the Macintosh, intubation force magnitude with Airtraq laryngoscopy was less dependent on glottic visualization.

CONCLUSIONS:

Previously, we reported that intubation force with the Airtraq was less in magnitude compared with the Macintosh. Our current study adds that intubation force also is less dependent on glottic view with Airtraq compared with the Macintosh.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Laringoscópios / Laringoscopia Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Laringoscópios / Laringoscopia Idioma: En Ano de publicação: 2022 Tipo de documento: Article