Your browser doesn't support javascript.
loading
The Utility of MRI With Multiacquisition Variable-Resonance Image Combination (MAVRIC) in Diagnosing Deep Total Hip Arthroplasty Infection.
Levack, Ashley E; Koch, Chelsea; Moore, Harold G; Cross, Michael B.
Afiliação
  • Levack AE; Department of Orthopaedic Surgery, Hospital for Special Surgery, New York, NY, USA.
  • Koch C; Department of Orthopaedic Surgery, Hospital for Special Surgery, New York, NY, USA.
  • Moore HG; Weill Cornell Medical College, New York, NY, USA.
  • Cross MB; Department of Orthopaedic Surgery, Hospital for Special Surgery, New York, NY, USA.
HSS J ; 18(2): 277-283, 2022 May.
Article em En | MEDLINE | ID: mdl-35645646
Background: The 2010 American Academy of Orthopaedic Surgeons Clinical Practice Guidelines report insufficient evidence to address the diagnostic efficacy of magnetic resonance imaging (MRI) for periprosthetic joint infection (PJI). Questions/Purposes: The purpose of this study was to determine the utility of MRI with multiacquisition variable-resonance image combination (MAVRIC) metal artifact suppression techniques in diagnosing PJI in the setting of total hip arthroplasty (THA). Methods: Multiacquisition variable-resonance image combination MRIs obtained of THAs between November 2012 and November 2016 were queried. Radiology reports were classified as positive (suspicious for infection), negative (no features of infection), or inconclusive (infection cannot be excluded or correlation with aspiration suggested if clinically concerned). Chart review identified cases of deep PJI according to the modified Musculoskeletal Infection Society criteria. Results: Of 2156 MRIs of THAs included, MRI was concerning for infection in 1.8% (n = 39), inconclusive in 1.2% (n = 26), and negative in 97.0% (n = 2091). Deep PJI was identified in 53 (2.5%) patients, 30 of whom (56.6%) had conclusively positive finding on MRI (false-negative rate: 43.4%, sensitivity: 56.6%). Of 2103 aseptic THAs, only 9 (0.4%) MRIs were read as suspicious for infection (false-positive rate: 0.4%; specificity: 99.6%). Conclusion: Magnetic resonance imaging with MAVRIC is a highly specific test for PJI with a low false-positive rate. This indicates that when clinicians are provided with an MRI that unexpectedly suggests infection, a formal evaluation for infection is indicated. In patients with otherwise equivocal diagnostic findings, MRI may help confirm, but not refute, a diagnosis of PJI. Prospective study with more experienced image reviewers may further support the use of MRI in PJI.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2022 Tipo de documento: Article