Multi-center study of inter-rater reproducibility, image quality, and diagnostic accuracy of CZT versus conventional SPECT myocardial perfusion imaging.
Mansour, Nabeel; Nekolla, Stephan G; Reyes, Eliana; Angelidis, George; Georgoulias, Panagiotis; Anagnostopoulos, Constantinos; Bravo, Paco; Bruno, Isabella; Flotats, Albert; Fuentes-Ocampo, Francisco; Sciagrà, Roberto; Keng, Felix; Kessler, Lukas; Papathanasiou, Maria; Soman, Prem; Rischpler, Christoph.
Afiliação
  • Mansour N; Department of Nuclear Medicine, Klinikum Rechts der Isar, Technical University Munich, Munich, Germany. Nabeel.Mansour@med.uni-muenchen.de.
  • Nekolla SG; Department of Radiology, University Hospital, LMU Munich, Marchioninistr. 15, 81377, Munich, Germany. Nabeel.Mansour@med.uni-muenchen.de.
  • Reyes E; Department of Nuclear Medicine, Klinikum Rechts der Isar, Technical University Munich, Munich, Germany.
  • Angelidis G; DZHK (German Centre for Cardiovascular Research), Partner Site Munich Heart Alliance, Munich, Germany.
  • Georgoulias P; Royal Brompton and Harefield NHS Trust, London, UK.
  • Anagnostopoulos C; Department of Nuclear Medicine, Faculty of Medicine, University of Thessaly, University Hospital of Larissa, Larissa, Greece.
  • Bravo P; Department of Nuclear Medicine, Faculty of Medicine, University of Thessaly, University Hospital of Larissa, Larissa, Greece.
  • Bruno I; BRFAA - Biomedical Research Foundation Academy of Athens, Athens, Greece.
  • Flotats A; Divisions of Nuclear Medicine and Cardiology, Departments of Radiology and Medicine, Hospital of the University of Pennsylvania, Philadelphia, PA, USA.
  • Fuentes-Ocampo F; Fondazione Policlinico Universitario Agostino Gemelli IRCCS, Roma, Italy.
  • Sciagrà R; Nuclear Medicine Department, Hospital de la Santa Creu i Sant Pau, Universitat Autònoma de Barcelona, Barcelona, Spain.
  • Keng F; Nuclear Medicine Department, Hospital de la Santa Creu i Sant Pau, Universitat Autònoma de Barcelona, Barcelona, Spain.
  • Kessler L; Department of Experimental and Clinical Biomedical Sciences 'Mario Serio', University of Florence, Florence, Italy.
  • Papathanasiou M; National Heart Centre, Singapore, Singapore.
  • Soman P; Department of Nuclear Medicine, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
  • Rischpler C; Department of Cardiology, West German Heart and Vascular Center, University Hospital, Essen, Germany.
J Nucl Cardiol ; 30(2): 528-539, 2023 04.
Article em En | MEDLINE | ID: mdl-35799039
ABSTRACT

BACKGROUND:

Cadmium-zinc-telluride (CZT)-based detectors exhibit higher diagnostic sensitivity in myocardial perfusion imaging (MPI) than conventional Anger-MPI for detection of coronary artery disease (CAD); however, reduced specificity and diagnostic accuracy of CZT-MPI were observed. This study aims to compare these different camera systems and to examine the degree of inter-rater reproducibility among readers with varying experience in MPI.

METHODS:

83 patients who underwent double stress/rest examinations using both a CZT and conventional SPECT cameras within one visit were included. Anonymized and randomized MPI-images were distributed to 15 international readers using a standardized questionnaire. Subsequent coronary angiography findings of ten patients served as a reference for analysis of sensitivity and specificity.

RESULTS:

Image quality was significantly better in CZT-MPI with significantly lower breast attenuation (P < 0.05). CZT-MPI exhibited higher sensitivity than Anger-MPI (87.5% vs. 62.5%) and significantly reduced specificity (40% vs. 100%). Readers experienced with both camera systems had the highest inter-rater agreement indicating higher reproducibility (CZT 0.54 vs. conv. 0.49, P < 0.05).

CONCLUSIONS:

Higher diagnostic sensitivity of CZT-MPI offers advantages in detection of CAD yet potentially of at the cost of reduced specificity, therefore it requires special training and a differentiated evaluation approach, especially for non-experienced readers with such camera systems.