Your browser doesn't support javascript.
loading
An Improved POD Model for Fast Semi-Quantitative Analysis of Carbendazim in Fruit by Surface Enhanced Raman Spectroscopy.
Yang, Qiaoling; Lin, Hong; Ma, Jinge; Chen, Niannian; Zhao, Chaomin; Guo, Dehua; Niu, Bing; Zhao, Zhihui; Deng, Xiaojun; Chen, Qin.
Afiliação
  • Yang Q; School of Environmental and Chemical Engineering, Shanghai University, Shanghai 200444, China.
  • Lin H; School of Life Sciences, Shanghai University, Shanghai 200444, China.
  • Ma J; Department of Endocrine and Metabolic Diseases, Shanghai Institute of Endocrine and Metabolic Diseases, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200025, China.
  • Chen N; Tech Ctr Anim Plant & Food Inspect & Quarantine, Shanghai Customs, Shanghai, 200135, China.
  • Zhao C; Tech Ctr Anim Plant & Food Inspect & Quarantine, Shanghai Customs, Shanghai, 200135, China.
  • Guo D; Tech Ctr Anim Plant & Food Inspect & Quarantine, Shanghai Customs, Shanghai, 200135, China.
  • Niu B; Tech Ctr Anim Plant & Food Inspect & Quarantine, Shanghai Customs, Shanghai, 200135, China.
  • Zhao Z; School of Life Sciences, Shanghai University, Shanghai 200444, China.
  • Deng X; Shanghai Oceanhood Instrument Equipment Co., Ltd., Shanghai 201608, China.
  • Chen Q; Tech Ctr Anim Plant & Food Inspect & Quarantine, Shanghai Customs, Shanghai, 200135, China.
Molecules ; 27(13)2022 Jun 30.
Article em En | MEDLINE | ID: mdl-35807472
ABSTRACT
The current detection method of carbendazim suffers from the disadvantages of complicated preprocessing and long cycle time. In order to solve the problem of rapid quantitative screening of finite contaminants, this article proposed a qualitative method based on characteristic peaks and a semi-quantitative method based on threshold to detect carbendazim in apple, and finally the method is evaluated by a validation system based on binary output. The results showed that the detection limit for carbendazim was 0.5 mg/kg, and the detection probability was 100% when the concentration was no less than 1 mg/kg. The semi-quantitative analysis method had a false positive rate of 0% and 5% at 0.5 mg/kg and 2.5 mg/kg, respectively. The results of method evaluation showed that when the added concentration was greater than 2.5 mg/kg, the qualitative detection method was consistent with the reference method. When the concentration was no less than 5 mg/kg, the semi-quantitative method is consistent between different labs. The semi-quantitative method proposed in this study can achieve the screening of finite contaminants in blind samples and simplify the test validation process through the detection probability model, which can meet the needs of rapid on-site detection and has a good application prospect.