Your browser doesn't support javascript.
loading
Intraarterial encephalography from an acutely implanted aneurysm embolization device in awake humans.
Li, Luyuan; Ibayashi, Kenji; Piscopo, Anthony; Deifelt Streese, Carolina; Chen, Haiming; Greenlee, Jeremy D W; Hasan, David M.
Afiliação
  • Li L; 1Department of Neurosurgery, University of Iowa, Iowa City, Iowa.
  • Ibayashi K; 2Department of Neurosurgery, Jichi Medical University, Tochigi, Japan; and.
  • Piscopo A; 1Department of Neurosurgery, University of Iowa, Iowa City, Iowa.
  • Deifelt Streese C; 1Department of Neurosurgery, University of Iowa, Iowa City, Iowa.
  • Chen H; 1Department of Neurosurgery, University of Iowa, Iowa City, Iowa.
  • Greenlee JDW; 1Department of Neurosurgery, University of Iowa, Iowa City, Iowa.
  • Hasan DM; 3Department of Neurosurgery, Duke University, Durham, North Carolina.
J Neurosurg ; 138(3): 785-792, 2023 03 01.
Article em En | MEDLINE | ID: mdl-35932270
ABSTRACT

OBJECTIVE:

Endovascular electroencephalography (evEEG) uses the cerebrovascular system to record electrical activity from adjacent neural structures. The safety, feasibility, and efficacy of using the Woven EndoBridge Aneurysm Embolization System (WEB) for evEEG has not been investigated.

METHODS:

Seventeen participants undergoing awake WEB endovascular treatment of unruptured cerebral aneurysms were included. After WEB deployment and before detachment, its distal deployment wire was connected to an EEG receiver, and participants performed a decision-making task for 10 minutes. WEB and scalp recordings were captured.

RESULTS:

All patients underwent successful embolization and evEEG with no complications. Event-related potentials were detected on scalp EEG in 9/17 (53%) patients. Of these 9 patients, a task-related low-gamma (30-70 Hz) response on WEB channels was captured in 8/9 (89%) cases. In these 8 patients, the WEB was deployed in 2 middle cerebral arteries, 3 anterior communicating arteries, the terminal internal carotid artery, and 2 basilar tip aneurysms. Electrocardiogram artifact on WEB channels was present in 12/17 cases.

CONCLUSIONS:

The WEB implanted within cerebral aneurysms of awake patients is capable of capturing task-specific brain electrical activities. Future studies are warranted to establish the efficacy of and support for evEEG as a tool for brain recording, brain stimulation, and brain-machine interface applications.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aneurisma Intracraniano / Embolização Terapêutica / Procedimentos Endovasculares Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aneurisma Intracraniano / Embolização Terapêutica / Procedimentos Endovasculares Idioma: En Ano de publicação: 2023 Tipo de documento: Article