Your browser doesn't support javascript.
loading
The Effects of Filtering PPG Signal on Pulse Arrival Time-Systolic Blood Pressure Correlation.
Annu Int Conf IEEE Eng Med Biol Soc ; 2022: 674-677, 2022 07.
Article em En | MEDLINE | ID: mdl-36086297
ABSTRACT
Pulse arrival time (PAT), evaluated from electro-cardiogram (ECG) and photoplethysmogram (PPG) signals, has been widely used for cuff-less blood pressure (BP) estimation due to its high correlation with BP. However, the question of whether filtering the PPG signal impacts the extracted PAT values and consequently, the correlation between PAT and BP, has not been investigated before. In this paper, using data from 18 subjects, changes in the PAT values, and in the subject-specific PAT-systolic BP (SBP) correlation caused by filtering the PPG signal with variable cutoff frequencies in the range of 2 to 15 Hz are studied. For PAT extraction, three PPG characteristic points (foot, maximum slope and systolic peak) are considered. Results show that differences in the cutoff frequency can shift the PAT values and introduce a worst-case error of over 8.2 mmHg for SBP estimation, indicating that PPG signal filter settings can impact PAT-based BP estimations. Our study suggests that extracting the PAT from the maximum slope point of PPG signal filtered at 10 Hz provides the most stable correlation with SBP.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Determinação da Pressão Arterial / Fotopletismografia Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Determinação da Pressão Arterial / Fotopletismografia Idioma: En Ano de publicação: 2022 Tipo de documento: Article