A novel and practical method to add video monitoring to tilt table testing.
de Lange, Frederik Jorrit; Hofland, Willem Petrus Merijn Emmanuël; Ferrara, Alessio; Gargaro, Alessio; Brignole, Michele; van Dijk, Jan Gerrit.
Afiliação
  • de Lange FJ; Amsterdam UMC, Department of Clinical and Experimental Cardiology, Amsterdam Cardiovascular Sciences, University of Amsterdam, Heart Centre, Meibergdreef 9, 1105AZ, Amsterdam, The Netherlands.
  • Hofland WPME; Amsterdam UMC, Department of Clinical and Experimental Cardiology, Amsterdam Cardiovascular Sciences, University of Amsterdam, Heart Centre, Meibergdreef 9, 1105AZ, Amsterdam, The Netherlands.
  • Ferrara A; BIOTRONIK Italia S.p.A., Clinical Research Unit, Via dell' Industria, 11, 20090 Milan, MI, Italy.
  • Gargaro A; BIOTRONIK Italia S.p.A., Clinical Research Unit, Via dell' Industria, 11, 20090 Milan, MI, Italy.
  • Brignole M; Department of Cardiovascular, Neural and Metabolic Sciences, Faint & Fall Programme, IRCCS Istituto Auxologico Italiano, San Luca Hospital, Piazzale Brescia, 20, 20149 Milan, MI, Italy.
  • van Dijk JG; Department of Neurology, Leiden University Medical Centre, Albinusdreef 2, Zuid Holland, 2333 ZA Leiden, The Netherlands.
Europace ; 25(2): 762-766, 2023 02 16.
Article em En | MEDLINE | ID: mdl-36351661
ABSTRACT

AIMS:

We describe a novel, practical, and inexpensive method to add video recording during tilt table testing (TTT) Open-Access-Video-TTT. METHODS AND

RESULTS:

The Open-Access-Video-TTT set-up uses a personal computer (PC) to capture screen video data from a non-invasive-beat-to-beat (NIBTB) haemodynamic blood pressure (BP) device, combined with video recording of a patient, using Open Broadcaster Software (OBS®). The new Open-Access-Video-TTT set up was tested with both the Finometer (model Finapres Nova®, Medical Systems, the Netherlands) and the Task Force® Touch Cardio monitor (CNSystems, Austria). For this, the Finapres Nova® was enabled in 'remote' mode and Real Video Network Computing (RealVNC®) was installed on the PC/laptop. The Task Force® has a DisplayPort (DP) port, for which a DP/ high-definition multimedia interface (HDMI) cable and a video capture card is used to merge the signals to the PC/laptop. With this method the combined images are stored as a new video signal. TTT can be performed with any routine protocol.

CONCLUSIONS:

Open Access-Video-TTT worked well for both the Finapres NOVA® and the Task Force Monitor ®. This novel method can be used easily by all physicians who wish to add video recording during TTT who do not have access to an electroencephalogram machine.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Teste da Mesa Inclinada / Hemodinâmica Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Teste da Mesa Inclinada / Hemodinâmica Idioma: En Ano de publicação: 2023 Tipo de documento: Article