Clinical application of the contrast-enhancement boost technique in computed tomography angiography of the portal vein.
Hou, Juan; Zhang, Yuan; Yan, Jing; Zhang, Tieliang; Xia, Wenwen; Zhu, Yucai; Liu, Wenya.
Afiliação
  • Hou J; Imaging Center, First Affiliated Hospital of Xinjiang Medical University, Urumqi, 830011, Xinjiang, China.
  • Zhang Y; Imaging Center, The Fourth Affiliated Hospital of Xinjiang Medical University, Urumqi, 830011, Xinjiang, China.
  • Yan J; Canon Medical Systems (China), Co., Ltd., Room 1004-1006, City Point, NO 666 West Huaihai Rd, Changning District, 200052, Shanghai, China.
  • Zhang T; Imaging Center, First Affiliated Hospital of Xinjiang Medical University, Urumqi, 830011, Xinjiang, China.
  • Xia W; Imaging Center, First Affiliated Hospital of Xinjiang Medical University, Urumqi, 830011, Xinjiang, China.
  • Zhu Y; Imaging Center, First Affiliated Hospital of Xinjiang Medical University, Urumqi, 830011, Xinjiang, China.
  • Liu W; Imaging Center, First Affiliated Hospital of Xinjiang Medical University, Urumqi, 830011, Xinjiang, China. 13999202977@163.com.
Abdom Radiol (NY) ; 48(2): 806-815, 2023 02.
Article em En | MEDLINE | ID: mdl-36471003
ABSTRACT

PURPOSE:

The aim of this study was to explore the improved image quality of the portal vein using the contrast-enhancement boost (CE-boost) technique for the improved visibility of abdominal-enhanced computed tomography (CT) scans in clinical practice.

METHODS:

This retrospective study included 50 patients in Group A who underwent routine abdominal-enhanced CT and 50 patients in Group B who underwent abdominal computed tomography angiography (CTA) with matched body mass index, age, and sex. Images in Group A were postprocessed with the CE-boost technique for further enhanced visibility of the portal vein. Both subjective and objective assessments of different branches of the portal vein in three types of images (i.e., Group A with CE-boost and without CE-boost, Group B) were statistically analyzed.

RESULTS:

The subjective scores of two experienced radiologists showed good consistency (kappa value > 0.624, p < 0.001), and the score of Group A with CE-boost (mean, 4.64) was significantly higher than that of the others (p < 0.001). The liver parenchyma and most target veins in Group A with CE-boost showed the highest CT, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) values and the lowest standard deviation (SD), while the CNR of most portal veins in Group A without CE-boost had the lowest CNR (p < 0.001). There were no differences in the SNR of the portal vein in Group A without CE-Boost and Group B (p > 0.05).

CONCLUSION:

CE-boost can significantly improve image quality in portal vein imaging without any additional scanning settings or changes in the clinical workflow.