Adjustment of Maternal Variable Rate Insulin Infusions Using Real-Time Continuous Glucose Monitoring in Pregnant Women with Type 1 Diabetes.
Diabetes Technol Ther
; 25(4): 293-297, 2023 04.
Article
em En
| MEDLINE
| ID: mdl-36695715
Real-time continuous glucose monitoring (CGM) in hospital holds promise; however, further evidence is required on its use to guide adjustment of variable rate intravenous insulin infusion (VRIII). We retrospectively analyzed data from 20 women with type 1 diabetes during the peripartum period who were commenced on VRIII. Data were analyzed for CGM accuracy (Dexcom G6) using point-of-care glucose-CGM matched pairs. The study was entirely observational, with no deviation from standard clinical care. Twenty women were included; median age 30 (26-35) years with first glycated hemoglobin in pregnancy of 57 (49-60) mmol/mol. Overall median absolute relative difference was 6.1 (1.6-17.3)%. The total simulated CGM-adjusted VRIII was 2.5 U per hour, compared with 2.4 U per hour with capillary blood glucose-adjusted VRIII. In this retrospective analysis of CGM adjustment of maternal VRIII, we demonstrate early feasibility and considerable accuracy. Further prospective studies are required to confirm the safety and potential efficacy of CGM-based insulin titration.