Your browser doesn't support javascript.
loading
Feedback lock-in: A versatile multi-terminal measurement system for electrical transport devices.
Barnard, Arthur W; Mikheev, Evgeny; Finney, Joe; Hiller, Han S; Goldhaber-Gordon, David.
Afiliação
  • Barnard AW; Department of Physics, Stanford University, Stanford, California 94305, USA.
  • Mikheev E; Department of Physics, Stanford University, Stanford, California 94305, USA.
  • Finney J; Department of Physics, Stanford University, Stanford, California 94305, USA.
  • Hiller HS; Department of Physics, University of Washington, Seattle, Washington 98195, USA.
  • Goldhaber-Gordon D; Department of Physics, Stanford University, Stanford, California 94305, USA.
Rev Sci Instrum ; 94(1): 013902, 2023 Jan 01.
Article em En | MEDLINE | ID: mdl-36725603
ABSTRACT
We present the design and implementation of a measurement system that enables parallel drive and detection of small currents and voltages at numerous electrical contacts to a multi-terminal electrical device. This system, which we term a feedback lock-in, combines digital control-loop feedback with software-defined lock-in measurements to dynamically source currents and measure small, pre-amplified potentials. The effective input impedance of each current/voltage probe can be set via software, permitting any given contact to behave as an open-circuit voltage lead or as a virtually grounded current source/sink. This enables programmatic switching of measurement configurations and permits measurement of currents at multiple drain contacts without the use of current preamplifiers. Our 32-channel implementation relies on commercially available digital input/output boards, home-built voltage preamplifiers, and custom open-source software. With our feedback lock-in, we demonstrate differential measurement sensitivity comparable to a widely used commercially available lock-in amplifier and perform efficient multi-terminal electrical transport measurements on twisted bilayer graphene and SrTiO3 quantum point contacts. The feedback lock-in also enables a new style of measurement using multiple current probes, which we demonstrate on a ballistic graphene device.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article