Design and performance simulation studies of a breast PET insert integrable into a clinical whole-body PET/MRI scanner.
Pommranz, C M; Schmidt, F P; Mannheim, J G; Diebold, S J; Tenzer, C; Santangelo, A; Pichler, B J.
Afiliação
  • Pommranz CM; Werner Siemens Imaging Center, Department of Preclinical Imaging and Radiopharmacy, Eberhard Karls University Tuebingen, Roentgenweg 13, D-72076 Tuebingen, Germany.
  • Schmidt FP; Institute for Astronomy and Astrophysics, Eberhard Karls University Tuebingen, Sand 1, D-72076 Tuebingen, Germany.
  • Mannheim JG; Werner Siemens Imaging Center, Department of Preclinical Imaging and Radiopharmacy, Eberhard Karls University Tuebingen, Roentgenweg 13, D-72076 Tuebingen, Germany.
  • Diebold SJ; Department of Nuclear Medicine and Clinical Molecular Imaging, University Hospital Tuebingen, Otfried-Mueller-Strasse 14, D-72076 Tuebingen, Germany.
  • Tenzer C; Werner Siemens Imaging Center, Department of Preclinical Imaging and Radiopharmacy, Eberhard Karls University Tuebingen, Roentgenweg 13, D-72076 Tuebingen, Germany.
  • Santangelo A; Cluster of Excellence iFIT (EXC 2180) Image Guided and Functionally Instructed Tumor Therapies, University of Tuebingen, Tuebingen, Germany.
  • Pichler BJ; Institute for Astronomy and Astrophysics, Eberhard Karls University Tuebingen, Sand 1, D-72076 Tuebingen, Germany.
Phys Med Biol ; 68(5)2023 02 28.
Article em En | MEDLINE | ID: mdl-36753773
ABSTRACT
Objective. Three different breast positron emission tomography (PET) insert geometries are proposed for integration into an existing magnetic resonance imaging (MRI) breast coil (Breast Biopsy Coil, NORAS MRI products) to be used inside a whole-body PET/MRI scanner (Biograph mMR, Siemens Healthineers) to enhance the sensitivity and spatial resolution of imaging inside the breast.Approach. Monte Carlo simulations were performed to predict and compare the performance characteristics of the three geometries in terms of the sensitivity, spatial resolution, scatter fraction, and noise equivalent count rate (NECR). In addition, the background single count rate due to organ uptake in a clinical scan scenario was predicted using a realistic anthropomorphic phantom.Main results. In the center of the field of view (cFOV), absolute sensitivities of 3.1%, 2.7%, and 2.2% were found for Geometry A (detectors arranged in two cylinders), Geometry B (detectors arranged in two partial cylinders), and Geometry C (detectors arranged in two half cylinders combined with two plates), respectively. The full width at half maximum spatial resolution was determined to be 1.7 mm (Geometry A), 1.8 mm (Geometry B) and 2.0 mm (Geometry C) at 5 mm from the cFOV. Designs with multiple scintillation-crystal layers capable of determining the depth of interaction (DOI) strongly improved the spatial resolution at larger distances from the transaxial cFOV. The system scatter fractions were 33.1% (Geometries A and B) and 32.3% (Geometry C). The peak NECRs occurred at source activities of 300 MBq (Geometry A), 310 MBq (Geometry B) and 340 MBq (Geometry C). The background single-event count rates were 17.1 × 106cps (Geometry A), 15.3 × 106cps (Geometry B) and 14.8 × 106cps (Geometry C). Geometry A in the three-layer DOI variant exhibited the best PET performance characteristics but could be challenging to manufacture. Geometry C had the lowest impact on the spatial resolution and the lowest sensitivity among the investigated geometries.Significance. Geometry B in the two-layer DOI variant represented an effective compromise between the PET performance and manufacturing difficulty and was found to be a promising candidate for the future breast PET insert.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Tomografia por Emissão de Pósitrons Idioma: En Ano de publicação: 2023 Tipo de documento: Article