Validation of a Machine Learning Expert Supporting System, ImmunoGenius, Using Immunohistochemistry Results of 3000 Patients with Lymphoid Neoplasms.
Diagnostics (Basel)
; 13(7)2023 Mar 31.
Article
em En
| MEDLINE
| ID: mdl-37046526
(1) Background: Differential diagnosis using immunohistochemistry (IHC) panels is a crucial step in the pathological diagnosis of hematolymphoid neoplasms. In this study, we evaluated the prediction accuracy of the ImmunoGenius software using nationwide data to validate its clinical utility. (2) Methods: We collected pathologically confirmed lymphoid neoplasms and their corresponding IHC results from 25 major university hospitals in Korea between 2015 and 2016. We tested ImmunoGenius using these real IHC panel data and compared the precision hit rate with previously reported diagnoses. (3) Results: We enrolled 3052 cases of lymphoid neoplasms with an average of 8.3 IHC results. The precision hit rate was 84.5% for these cases, whereas it was 95.0% for 984 in-house cases. (4) Discussion: ImmunoGenius showed excellent results in most B-cell lymphomas and generally showed equivalent performance in T-cell lymphomas. The primary reasons for inaccurate precision were atypical IHC profiles of certain cases, lack of disease-specific markers, and overlapping IHC profiles of similar diseases. We verified that the machine-learning algorithm could be applied for diagnosis precision with a generally acceptable hit rate in a nationwide dataset. Clinical and histological features should also be taken into account for the proper use of this system in the decision-making process.
Texto completo:
1
Base de dados:
MEDLINE
Idioma:
En
Ano de publicação:
2023
Tipo de documento:
Article