A robust motion correction technique for infrared thermography during awake craniotomy.
Iorga, Michael; Tate, Matthew C; Parrish, Todd B.
Afiliação
  • Iorga M; Department of Radiology, Northwestern University, Chicago, IL, USA. michael.iorga@northwestern.edu.
  • Tate MC; Department of Biomedical Engineering, Northwestern University, Evanston, IL, USA. michael.iorga@northwestern.edu.
  • Parrish TB; Department of Neurosurgery, Northwestern Medicine, Chicago, IL, USA.
Int J Comput Assist Radiol Surg ; 18(12): 2223-2231, 2023 Dec.
Article em En | MEDLINE | ID: mdl-37222929
ABSTRACT

PURPOSE:

Intraoperative infrared thermography is an emerging technique for image-guided neurosurgery, whereby physiological and pathological processes result in temperature changes over space and time. However, motion during data collection leads to downstream artifacts in thermography analyses. We develop a fast, robust technique for motion estimation and correction as a preprocessing step for brain surface thermography recordings.

METHODS:

A motion correction technique for thermography was developed which approximates the deformation field associated with motion as a grid of two-dimensional bilinear splines (Bispline registration), and a regularization function was designed to constrain motion to biomechanically feasible solutions. The performance of the proposed Bispline registration technique was compared to phase correlation, a band-stop filter, demons registration, and the Horn-Schunck and Lucas-Kanade optical flow techniques.

RESULTS:

All methods were analyzed using thermography data from ten patients undergoing awake craniotomy for brain tumor resection, and performance was compared using image quality metrics. The proposed method had the lowest mean-squared error and the highest peak-signal-to-noise ratio of all methods tested and performed slightly worse than phase correlation and Demons registration on the structural similarity index metric (p < 0.01, Wilcoxon signed-rank test). Band-stop filtering and the Lucas-Kanade method were not strong attenuators of motion, while the Horn-Schunck method was well-performing initially but weakened over time.

CONCLUSION:

Bispline registration had the most consistently strong performance out of all the techniques tested. It is relatively fast for a nonrigid motion correction technique, capable of processing ten frames per second, and could be a viable option for real-time use. Constraining the deformation cost function through regularization and interpolation appears sufficient for fast, monomodal motion correction of thermal data during awake craniotomy.