Diamond-like carbon coating to inner surface of polyurethane tube reduces Staphylococcus aureus bacterial adhesion and biofilm formation.
J Artif Organs
; 2023 May 25.
Article
em En
| MEDLINE
| ID: mdl-37227545
Staphylococcus aureus is one of the main causative bacteria for polyurethane catheter and artificial graft infection. Recently, we developed a unique technique for coating diamond-like carbon (DLC) inside the luminal resin structure of polyurethane tubes. This study aimed to elucidate the infection-preventing effects of diamond-like carbon (DLC) coating on a polyurethane surface against S. aureus. We applied DLC to polyurethane tubes and rolled polyurethane sheets with our newly developed DLC coating technique for resin tubes. The DLC-coated and uncoated polyurethane surfaces were tested in smoothness, hydrophilicity, zeta-potential, and anti-bacterial properties against S. aureus (biofilm formation and bacterial attachment) by contact with bacterial fluids under static and flow conditions. The DLC-coated polyurethane surface was significantly smoother, more hydrophilic, and had a more negative zeta-potential than did the uncoated polyurethane surface. Upon exposure to bacterial fluid under both static and flow conditions, DLC-coated polyurethane exhibited significantly less biofilm formation than uncoated polyurethane, based on absorbance measurements. In addition, the adherence of S. aureus was significantly lower for DLC-coated polyurethane than for uncoated polyurethane under both conditions, based on scanning electron microscopy. These results show that applying DLC coating to the luminal resin of polyurethane tubes may impart antimicrobial effects against S. aureus to implantable medical polyurethane devices, such as vascular grafts and central venous catheters.
Texto completo:
1
Base de dados:
MEDLINE
Idioma:
En
Ano de publicação:
2023
Tipo de documento:
Article