Impact of shield location on staff and caregiver dose rates for I-131 radiopharmaceutical therapy patients.
Bellamy, M; Chu, B; Serencsits, B; Quinn, B; Prasad, K; Altamirano, J; Williamson, M; Miodownik, D; Abrahams, N; Chen, F; Bierman, D; Wutkowski, M; Carter, L; Dauer, L.
Afiliação
  • Bellamy M; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Chu B; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Serencsits B; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Quinn B; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Prasad K; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Altamirano J; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Williamson M; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Miodownik D; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Abrahams N; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Chen F; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Bierman D; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Wutkowski M; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Carter L; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
  • Dauer L; Memorial Sloan Kettering, 1275 York Avenue, New York, NY 10065, United States of America.
J Radiol Prot ; 43(3)2023 07 21.
Article em En | MEDLINE | ID: mdl-37413983
ABSTRACT
The goal of this study is to investigate the effect of the location and width of a single lead shield on the dose rate of staff and caregivers in a hospital room with an I-131 patient. The best orientation of the patient and caregiver relative to the shield was determined based on minimizing staff and caregiver radiation dose rates. Shielded and unshielded dose rates were simulated using a Monte Carlo computer simulation and validated using real-world ionisation chamber measurements. Based on a radiation transport analysis using an adult voxel phantom published by the International Commission on Radiological Protection, placing the shield near the caregiver yielded the lowest dose rates. However, this strategy reduced the dose rate in only a tiny area of the room. Furthermore, positioning the shield near the patient in the caudal direction provided a modest dose rate reduction while shielding a large room area. Finally, increased shield width was associated with decreasing dose rates, but only a four-fold dose-rate reduction was observed for standard width shields. The recommendations of this case study may be considered as potential candidate room configurations where radiation dose rates are minimized, however these findings must be weighed against additional clinical, safety, and comfort considerations.