Your browser doesn't support javascript.
loading
Mid-term clinical and echocardiographic results of the INSPIRIS RESILIA aortic valve: a retrospective comparison to the Magna Ease.
Bernard, Jérémy; Georges, Gabriel; Hecht, Sébastien; Pibarot, Philippe; Clavel, Marie-Annick; Babaki, Shervin; Kalavrouziotis, Dimitri; Mohammadi, Siamak.
Afiliação
  • Bernard J; Cardiology Division, Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval/Quebec Heart and Lung Institute - Laval University, Quebec, QC, Canada.
  • Georges G; Cardiac Surgery Division, Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval/Quebec Heart and Lung Institute - Laval University, Quebec, QC, Canada.
  • Hecht S; Cardiology Division, Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval/Quebec Heart and Lung Institute - Laval University, Quebec, QC, Canada.
  • Pibarot P; Cardiology Division, Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval/Quebec Heart and Lung Institute - Laval University, Quebec, QC, Canada.
  • Clavel MA; Cardiology Division, Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval/Quebec Heart and Lung Institute - Laval University, Quebec, QC, Canada.
  • Babaki S; Research Division, Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval/Quebec Heart and Lung Institute - Laval University, Quebec, QC, Canada.
  • Kalavrouziotis D; Cardiac Surgery Division, Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval/Quebec Heart and Lung Institute - Laval University, Quebec, QC, Canada.
  • Mohammadi S; Cardiac Surgery Division, Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval/Quebec Heart and Lung Institute - Laval University, Quebec, QC, Canada.
Article em En | MEDLINE | ID: mdl-37462610
OBJECTIVES: The INSPIRIS aortic valve combines the RESILIA proprietary tissue preservation process and an expandable stent frame to benefit future transcatheter valve-in-valve procedures. As the INSPIRIS valve became commercially available in 2017, mid-term outcome reports are scarce. We aimed to evaluate mid-term safety and echocardiographic performance of the INSPIRIS valve in comparison to its predecessor, the Carpentier Edwards Perimount Magna Ease (ME). METHODS: This study was a retrospective single-centre study. Clinical results included early postoperative outcomes, mid-term mortality and readmission for cardiovascular cause or stroke. Echocardiographic follow-up (FU) was performed at discharge and 1-3, 6, 12 and 24 months. Clinical end point analyses were accomplished with a propensity score matching analysis and FU echocardiographic data comparisons using pairwise analyses and linear mixed-effect models. RESULTS: We included 953 patients who received an INSPIRIS (n = 488) or ME (n = 463) bioprosthesis between January 2018 and July 2021. In the matched population (n = 217 per group), no significant difference in short-term outcomes was observed, survival was similar at 30 months (INSPIRIS: 94% vs ME: 91%, P = 0.89), but freedom from readmission was higher in the INSPIRIS group (94% vs 86%, P = 0.014). INSPIRIS valves had a lower gradient at discharge (∼10 vs 14 mmHg, P < 0.001), 1-3 months (∼10 vs 12 mmHg, P < 0.001) and 24 months (∼11 vs 17 mmHg, P < 0.001) in paired analyses and significantly lower evolution of mean transvalvular gradients compared to ME. CONCLUSIONS: This study represents the largest comparative evaluation of the INSPIRIS to the ME valves, which demonstrated safe clinical outcomes and favourable haemodynamic performance at 2 years. Long-term FU is underway.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article