An integrated microfluidic-based biosensor using a magnetically controlled MNPs-enzyme microreactor to determine cholesterol in serum with fluorometric detection.
Román-Pizarro, Vanesa; Écija-Arenas, Ángela; Fernández-Romero, Juan M.
Afiliação
  • Román-Pizarro V; Departamento de Química Analítica, Instituto Universitario de Investigación en Química Fina Y Nanoquímica (IUNAN), Universidad de Córdoba, Campus de Rabanales, "Marie Curie" Building Annex, 14071, Córdoba, Spain.
  • Écija-Arenas Á; Departamento de Química Analítica, Instituto Universitario de Investigación en Química Fina Y Nanoquímica (IUNAN), Universidad de Córdoba, Campus de Rabanales, "Marie Curie" Building Annex, 14071, Córdoba, Spain.
  • Fernández-Romero JM; Departamento de Química Analítica, Instituto Universitario de Investigación en Química Fina Y Nanoquímica (IUNAN), Universidad de Córdoba, Campus de Rabanales, "Marie Curie" Building Annex, 14071, Córdoba, Spain. qa1feroj@uco.es.
Mikrochim Acta ; 190(8): 303, 2023 07 19.
Article em En | MEDLINE | ID: mdl-37464062
ABSTRACT
This work provides a microfluidic-based biosensor to determine total cholesterol in serum based on integrating the reaction/detection zone of a microfluidic chip of a magnetically retained enzyme microreactor (MREµR) coupled with the remote fluorometric detection through a bifurcated fiber-optic bundle (BFOB) connected with a conventional spectrofluorometer. The method is based on developing the enzymatic hydrolysis and oxidation of cholesterol at microscale size using both enzymes (cholesterol esterase (ChE) and cholesterol oxidase (ChOx)) immobilized on magnetic nanoparticles (MNPs). The biocatalyst reactions were followed by monitoring the fluorescence decreasing by the naphtofluorescein (NF) oxidation in the presence of the previous H2O2 formed. This microfluidic biosensor supposes the physical integration of a minimal MREµR as a bioactive enzyme area and the focused BFOB connected with the spectrofluorometer detector. The MREµR was formed by a 1 mm length of magnetic retained 21 ChE-MNP/ChOx-MNP mixture. The dynamic range of the calibration graph was 0.005-10 mmol L-1, expressed as total cholesterol concentration with a detection limit of 1.1 µmol L-1 (r2 = 0.9999, sy/x = 0.03, n = 10, r = 3). The precision expressed as the relative standard deviation (RSD%) was between 1.3 and 2.1%. The microfluidic-based biosensors showed a sampling frequency estimated at 30 h-1. The method was applied to determine cholesterol in serum samples with recovery values between 94.8 and 102%. The results of the cholesterol determination in serum were also tested by correlation with those obtained using the other two previous methods.