Your browser doesn't support javascript.
loading
DynamX sirolimus-eluting Bioadaptor versus the zotarolimus-eluting Resolute Onyx stent in patients with de novo coronary artery lesions: Design and rationale of the multi-center, international, randomized BIODAPTOR-RCT.
Saito, Shigeru; Nef, Holger M; Webster, Mark; Verheye, Stefan.
Afiliação
  • Saito S; Iryohojin Tokushukai Shonan Kamakura General Hospital, Japan. Electronic address: transradial@kamakuraheart.org.
  • Nef HM; Department of Cardiology, University of Giessen, Germany. Electronic address: Holger.Nef@innere.med.uni-giessen.de.
  • Webster M; Department of Cardiology, Auckland City Hospital, Auckland, New Zealand. Electronic address: MWebster@adhb.govt.nz.
  • Verheye S; Interventional Cardiology, ZNA Cardiovascular Center Middelheim, Antwerp, Belgium.
Cardiovasc Revasc Med ; 55: 76-82, 2023 10.
Article em En | MEDLINE | ID: mdl-37479544
ABSTRACT

BACKGROUND:

Conventional drug-eluting stents achieve good safety and performance outcomes, but the stents permanently cage the vessel, leading to a non-plateauing rate of clinical events. The DynamX Bioadaptor is designed to reduce these long-term events through unique design features that permit restoring vessel function and physiology through the disengagement of uncaging elements after the resorption of a biodegradable polymer over six months. Promising initial results have been obtained in the DynamX mechanistic study, with excellent safety and effectiveness, positive arterial remodeling, improved vasomotion, compliance, and cyclic pulsatility. We now aim to confirm these findings randomizing the DynamX Bioadaptor against the Resolute Onyx stent.

METHODS:

This multi-center, international, randomized single-blinded study is conducted in 34 sites across Europe, Japan, and New Zealand and is divided into the European/New Zealand cohort and the Japanese cohort (which includes an imaging subset). It is designed to randomly assign 444 patients (222 per region) in a 11 ratio to either the DynamX Bioadaptor or the Resolute Onyx stent. Furthermore, a pharmacokinetic substudy is conducted in 9 patients enrolled in Japan to assess the pharmacokinetics of sirolimus after implantation of the DynamX Bioadaptor. Study follow-up is scheduled at one, six, and 12 months, and annually thereafter for five years; imaging follow-up includes angiographic, intravascular ultrasound, and optical coherence tomography assessments at 12 months in a subset of patients. The primary endpoint is 12-month target lesion failure.

CONCLUSIONS:

This trial will provide valuable insights into the safety and efficacy of this novel bioadaptor when compared to a contemporary drug-eluting stent. CONDENSED ABSTRACT The DynamX Sirolimus-Eluting Bioadaptor has unique design features aiming to reduce long-term events after percutaneous coronary intervention by permitting the restoration of vessel function through the freeing of uncaging elements. Promising initial results have been obtained in the DynamX mechanistic study. This trial aims to confirm these findings in a randomized setting. The European/ New Zealand and Japanese cohorts were designed to randomly assign 444 subjects in a 11 ratio to either the DynamX Bioadaptor or the Resolute Onyx stent. Furthermore, a pharmacokinetic substudy is conducted in 9 patients enrolled in Japan to assess the pharmacokinetics of sirolimus.