Dynamic torque and screw-in force of four different glide path instruments assessed in simulated single- and double-curved canals: An in vitro study.
Thu, Myint; Ebihara, Arata; Maki, Keiichiro; Kimura, Shunsuke; Kyaw, Moe-Sandar; Kasuga, Yuka; Nishijo, Miki; Okiji, Takashi.
Afiliação
  • Thu M; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Ebihara A; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Maki K; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Kimura S; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Kyaw MS; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Kasuga Y; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Nishijo M; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Okiji T; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
J Dent Sci ; 18(4): 1598-1603, 2023 Oct.
Article em En | MEDLINE | ID: mdl-37799866
ABSTRACT
Background/

purpose:

The glide path instruments are the introductory instruments into the canals; thus, they should be durable enough to withstand torsional stress/screw-in force. This study aimed to investigate the torque and screw-in force of TruNatomy Glider (TN), ProGLIDER (PG), Hyflex EDM (EDM) and Dent Craft RE (RE) glide path instruments in single- and double-curved canals. Materials and

methods:

Each instrument brand was divided into two groups (n = 7 each) according to the canal configuration. Torque and screw-in force were recorded during automated instrumentation of simulated resin canals with XSmart IQ cordless motor after the canal patency was checked with a #10 K-file. The values were statistically analyzed using the Kolmogorov-Smirnov test followed by the Kruskal Wallis test and the Mann-Whitney U test with Bonferroni correction (α = 0.05).

Results:

TN produced significantly higher torque than RE in single-curved canals and PG in double-curved canals (P < 0.05). EDM yielded significantly higher screw-in force than TN and RE in single-curved canals (P < 0.05), but there was no significant difference in double-curved canals (P > 0.05). A significant effect of different canal configurations was only detected for screw-in force in EDM (P < 0.05).

Conclusion:

TN in single-curved canals and RE in double-curved canals yielded higher torque values, while EDM exhibited greater screw-in force in both canal configurations. No effect of different canal configurations was detected for torque, but a significant impact was detected for screw-in force in EDM.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article