Dual Suture Versus Suture and Plug Closure Devices for Large Bore Access Haemostasis During Percutaneous Access Endovascular Aneurysm Repair.
Cardiovasc Intervent Radiol
; 46(11): 1553-1561, 2023 Nov.
Article
em En
| MEDLINE
| ID: mdl-37828234
ABSTRACT
PURPOSE:
This study aimed to compare a dual Proglide strategy versus a combination of one Proglide and dual Exoseal for large-bore access closure during percutaneous access endovascular aneurysm repair (pEVAR). MATERIALS ANDMETHODS:
We retrospectively analyzed 97 patients who underwent pEVAR at our center between January 2021 and February 2023. The patients were divided into two groups dual Proglide (P + P) and one Proglide with dual Exoseal (P + E). The primary outcome measures were technical success and access-related vascular complications. Technical success was defined as achieving complete hemostasis without a bailout strategy. Postprocedural follow-up for access-related vascular complications was evaluated at 30 and 60 days using computed tomography angiography and ultrasonography. Severity was graded according to the Cardiovascular Interventional Radiological Society of Europe (CIRSE) Classification.RESULTS:
Overall, a dual Proglide strategy was used in 46 patients (47.4%) with 65 groins (46.4%), and a combination of one Proglide and dual Exoseal was used in 51 patients (52.6%) with 75 groins (53.6%). The baseline characteristics were similar between the groups. The total technical success rate was 96.4%, and no significant differences were observed (95.4% vs. 97.3%; p = 0.870). Minor bleeding treatable through compression occurred significantly more often in the P group (CIRSE 1, 10.8% vs. 1.3%, p = 0.042). Hemostasis time, procedural time, length of stay in the hospital, closure device failure, and incidence of unplanned intervention did not differ significantly between the groups.CONCLUSIONS:
A combined Proglide and Exoseal strategy is safe and effective for large-bore access closure during pEVAR and can be considered an alternative. However, it should be supported by larger prospective studies.Palavras-chave