Your browser doesn't support javascript.
loading
Computational Comparison of the Mechanical Behavior of Aortic Stent-Grafts Derived from Auxetic Unit Cells.
Vellaparambil, Rahul; Han, Woo-Suck; Di Giovanni, Pierluigi; Avril, Stéphane.
Afiliação
  • Vellaparambil R; Mines Saint-Etienne, Université Jean Monnet Saint-Etienne, INSERM, SAINBIOSE U1059, 42023, Saint-Etienne, France.
  • Han WS; Research and Development Department, HSL S.R.L, Trento, Italy.
  • Di Giovanni P; Mines Saint-Etienne, Université Jean Monnet Saint-Etienne, INSERM, SAINBIOSE U1059, 42023, Saint-Etienne, France.
  • Avril S; Research and Development Department, HSL S.R.L, Trento, Italy.
Cardiovasc Eng Technol ; 15(2): 199-210, 2024 Apr.
Article em En | MEDLINE | ID: mdl-38110763
ABSTRACT

PURPOSE:

Inappropriate stent-graft (SG) flexibility has been frequently associated with endovascular aortic repair (EVAR) complications such as endoleaks, kinks, and SG migration, especially in tortuous arteries. Stents derived from auxetic unit cells have shown some potential to address these issues as they offer an optimum trade-off between radial stiffness and bending flexibility.

METHODS:

In this study, we utilized an established finite element (FE)-based approach to replicate the mechanical response of a SG iliac limb derived from auxetic unit cells in a virtual tortuous iliac aneurysm using a combination of a 180° U-bend and intraluminal pressurization. This study aimed to compare the mechanical performance (flexibility and durability) of SG limbs derived from auxetic unit cells and two commercial SG limbs (Z-stented SG and circular-stented SG models) in a virtual tortuous iliac aneurysm. Maximal graft strain and maximum stress in stents were employed as criteria to estimate the durability of SGs, whereas the maximal luminal reduction rate and the bending stiffness were used to assess the flexibility of the SGs.

RESULTS:

SG limbs derived from auxetic unit cells demonstrated low luminal reduction (range 4-12%) with no kink, in contrast to Z-stented SG, which had a kink in its central area alongside a high luminal reduction (44%).

CONCLUSIONS:

SG limbs derived from auxetic unit cells show great promise for EVAR applications even at high angulations such as 180°, with acceptable levels of durability and flexibility.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Estresse Mecânico / Prótese Vascular / Stents / Implante de Prótese Vascular / Análise de Elementos Finitos / Procedimentos Endovasculares / Modelos Cardiovasculares Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Estresse Mecânico / Prótese Vascular / Stents / Implante de Prótese Vascular / Análise de Elementos Finitos / Procedimentos Endovasculares / Modelos Cardiovasculares Idioma: En Ano de publicação: 2024 Tipo de documento: Article