Your browser doesn't support javascript.
loading
Successful Bail Out of Rotablator Driveshaft Fracture Due to Severe Calcified Lesion and Proximal Tortuosity.
Uehara, Hiroki; Okuyama, Masaki; Oe, Yutaro; Yoshimura, Takaki; Gunji, Takahiro.
Afiliação
  • Uehara H; Department of Cardiovascular Medicine, Kin-Ikyo Chuo Hospital, Sapporo, Japan.
  • Okuyama M; Department of Cardiovascular Medicine, Kin-Ikyo Chuo Hospital, Sapporo, Japan.
  • Oe Y; Department of Cardiovascular Medicine, Kin-Ikyo Chuo Hospital, Sapporo, Japan.
  • Yoshimura T; Department of Cardiovascular Medicine, Kin-Ikyo Chuo Hospital, Sapporo, Japan.
  • Gunji T; Department of Cardiovascular Medicine, Kin-Ikyo Chuo Hospital, Sapporo, Japan.
Heart Views ; 24(4): 221-223, 2023.
Article em En | MEDLINE | ID: mdl-38188706
ABSTRACT
A 59-year-old male dialysis patient with a history of coronary artery bypass graft surgery underwent percutaneous coronary intervention of a right coronary artery with a severely calcified lesion. While debulking calcification using a Rotablator, the driveshaft was suddenly fractured. We attempted to move a child-in-mother catheter closer and across a second floppy wire and we dilated a 2.0-mm noncompliant balloon to trap the driveshaft, which we then removed, and restarted the intervention.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article