Finite element analysis and clinical application of 3D-printed Ti alloy implant for the reconstruction of mandibular defects.
BMC Oral Health
; 24(1): 95, 2024 Jan 17.
Article
em En
| MEDLINE
| ID: mdl-38233785
ABSTRACT
BACKGROUND:
The reconstruction of segmental defect of the mandible has always been a challenge. The customized reconstruction plate without a bone graft is also considered a transitional means of rehabilitation and reconstruction in some cases.METHODS:
This study evaluated the biomechanical behaviors of customized plates with different structural designs comparing with commercial plates using the finite element method in reconstrution of the lateral mandible defect.RESULTS:
Simulations revealed the stress state in the plate bodies, bone tissues and screws were associated with the width, height, thickness of the plates as well as the distribution of screws. In all of the groups, the system of 16 mm-high, 2.8 mm-thick customized reconstruction plate with 10 screws was considered to be the most ideal design because of the most harmonious biomechanical state. What's more, the stress shielding effects were not obvious in this experiment. Based on the above findings, we conducted a clinical case analysis to verify the mechanical properties of customized reconstruction and obtained a satisfactory operation result.CONCLUSIONS:
The results show that by adjusting the contour parameters of the reconstruction plates, an ideal and reliable customized plate can be manufactured. And the customized 3D-printed Ti alloy implant will be a new way to achieve mandibular reconstruction in patients unable to perform autologous bone graft surgery. TRIAL REGISTRATION The present trial has been registered with ChiCTR, the registration number is ChiCTR 2,000,038,973 on 11/10/2020.Palavras-chave