An LC-MS/MS method for serum cystatin C quantification and its comparison with two commercial immunoassays.
Zhang, Li; Nizhamuding, Xiaerbanu; Zheng, Hao; Zeng, Jie; Yuan, Xinyi; Ma, Zijia; Zhou, Weiyan; Zhang, Chao; Zhang, Tianjiao; Zhang, Chuanbao.
Afiliação
  • Zhang L; National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, Beijing, P.R. China.
  • Nizhamuding X; National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, P.R. China.
  • Zheng H; National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, Beijing, P.R. China.
  • Zeng J; National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, P.R. China.
  • Yuan X; National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, Beijing, P.R. China.
  • Ma Z; National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Beijing, P.R. China.
  • Zhou W; National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, Beijing, P.R. China.
  • Zhang C; National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Beijing, P.R. China.
  • Zhang T; National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, Beijing, P.R. China.
  • Zhang C; National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, P.R. China.
Clin Chem Lab Med ; 62(6): 1092-1100, 2024 May 27.
Article em En | MEDLINE | ID: mdl-38253403
ABSTRACT

OBJECTIVES:

The standardization of cystatin C (CysC) measurement has received increasing attention in recent years due to its importance in estimating glomerular filtration rate (GFR). Mass spectrometry-based assays have the potential to provide an accuracy base for CysC measurement. However, a precise, accurate and sustainable LC-MS/MS method for CysC is still lacking.

METHODS:

The developed LC-MS/MS method quantified CysC by detecting signature peptide (T3) obtained from tryptic digestion. Stable isotope labeled T3 peptide (SIL-T3) was spiked to control matrix effects and errors caused by liquid handling. The protein denaturation, reduction and alkylation procedures were combined into a single step with incubation time of 1 h, and the digestion lasted for 3.5 h. In the method validation, digestion time-course, imprecision, accuracy, matrix effect, interference, limit of quantification (LOQ), carryover, linearity, and the comparability to two routine immunoassays were evaluated.

RESULTS:

No significant matrix effect or interference was observed with the CysC measurement. The LOQ was 0.21 mg/L; the within-run and total imprecision were 1.33-2.05 % and 2.18-3.90 % for three serum pools (1.18-5.34 mg/L). The LC-MS/MS method was calibrated by ERM-DA471/IFCC and showed good correlation with two immunoassays traceable to ERM-DA471/IFCC. However, significant bias was observed for immunoassays against the LC-MS/MS method.

CONCLUSIONS:

The developed LC-MS/MS method is robust and simpler and holds the promise to provide an accuracy base for routine immunoassays, which will promote the standardization of CysC measurement.