Your browser doesn't support javascript.
loading
Evaluating false-positive detection in a computer-aided detection system for colonoscopy.
Okumura, Taishi; Imai, Kenichiro; Misawa, Masashi; Kudo, Shin-Ei; Hotta, Kinichi; Ito, Sayo; Kishida, Yoshihiro; Takada, Kazunori; Kawata, Noboru; Maeda, Yuki; Yoshida, Masao; Yamamoto, Yoichi; Minamide, Tatsunori; Ishiwatari, Hirotoshi; Sato, Junya; Matsubayashi, Hiroyuki; Ono, Hiroyuki.
Afiliação
  • Okumura T; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Imai K; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Misawa M; Digestive Disease Center, Showa University Northern Yokohama Hospital, Yokohama, Japan.
  • Kudo SE; Digestive Disease Center, Showa University Northern Yokohama Hospital, Yokohama, Japan.
  • Hotta K; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Ito S; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Kishida Y; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Takada K; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Kawata N; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Maeda Y; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Yoshida M; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Yamamoto Y; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Minamide T; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Ishiwatari H; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Sato J; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Matsubayashi H; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
  • Ono H; Division of Endoscopy, Shizuoka Cancer Center, Shizuoka, Japan.
J Gastroenterol Hepatol ; 39(5): 927-934, 2024 May.
Article em En | MEDLINE | ID: mdl-38273460
ABSTRACT
BACKGROUND AND

AIM:

Computer-aided detection (CADe) systems can efficiently detect polyps during colonoscopy. However, false-positive (FP) activation is a major limitation of CADe. We aimed to compare the rate and causes of FP using CADe before and after an update designed to reduce FP.

METHODS:

We analyzed CADe-assisted colonoscopy videos recorded between July 2022 and October 2022. The number and causes of FPs and excessive time spent by the endoscopist on FP (ET) were compared pre- and post-update using 11 propensity score matching.

RESULTS:

During the study period, 191 colonoscopy videos (94 and 97 in the pre- and post-update groups, respectively) were recorded. Propensity score matching resulted in 146 videos (73 in each group). The mean number of FPs and median ET per colonoscopy were significantly lower in the post-update group than those in the pre-update group (4.2 ± 3.7 vs 18.1 ± 11.1; P < 0.001 and 0 vs 16 s; P < 0.001, respectively). Mucosal tags, bubbles, and folds had the strongest association with decreased FP post-update (pre-update vs post-update 4.3 ± 3.6 vs 0.4 ± 0.8, 0.32 ± 0.70 vs 0.04 ± 0.20, and 8.6 ± 6.7 vs 1.6 ± 1.7, respectively). There was no significant decrease in the true positive rate (post-update vs pre-update 95.0% vs 99.2%; P = 0.09) or the adenoma detection rate (post-update vs pre-update 52.1% vs 49.3%; P = 0.87).

CONCLUSIONS:

The updated CADe can reduce FP without impairing polyp detection. A reduction in FP may help relieve the burden on endoscopists.