Your browser doesn't support javascript.
loading
Ultrashort VIBE-Dixon versus Cine and late gadolinium enhancement MRI sequences for cardiac thrombus detection.
Soschynski, Martin; Capilli, Fabio; Taha Hagar, Muhammad; Ruile, Philipp; Hein, Manuel; Benndorf, Matthias; Taron, Jana; Schlett, Christopher L; Bamberg, Fabian; Krauss, Tobias.
Afiliação
  • Soschynski M; Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Hugstetter Straße 55, 79106 Freiburg, Germany. Electronic address: Martin.Soschynski@uniklinik-freiburg.de.
  • Capilli F; Department of Radiology, Neuroradiology and Nuclear Medicine, Medical Center Vest, Ruhr University Bochum, Recklinghausen, Dorstener Straße 151, 45657 Recklinghausen, Germany. Electronic address: fabio.capilli@klinikum-vest.de.
  • Taha Hagar M; Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Hugstetter Straße 55, 79106 Freiburg, Germany. Electronic address: taha.hagar@uniklinik-freiburg.de.
  • Ruile P; Department of Cardiology and Angiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Hugstetter Straße 55, 79106 Freiburg, Germany. Electronic address: philipp.ruile@uniklinik-freiburg.de.
  • Hein M; Department of Cardiology and Angiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Hugstetter Straße 55, 79106 Freiburg, Germany. Electronic address: manuel.hein@uniklinik-freiburg.de.
  • Benndorf M; Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Hugstetter Straße 55, 79106 Freiburg, Germany. Electronic address: matthias.benndorf@uniklinik-freiburg.de.
  • Taron J; Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Hugstetter Straße 55, 79106 Freiburg, Germany. Electronic address: jana.taron@uniklinik-freiburg.de.
  • Schlett CL; Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Hugstetter Straße 55, 79106 Freiburg, Germany. Electronic address: christopher.schlett@uniklinik-freiburg.de.
  • Bamberg F; Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Hugstetter Straße 55, 79106 Freiburg, Germany. Electronic address: fabian.bamberg@uniklinik-freiburg.de.
  • Krauss T; Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Hugstetter Straße 55, 79106 Freiburg, Germany. Electronic address: tobias.krauss@uniklinik-freiburg.de.
Eur J Radiol ; 173: 111360, 2024 Apr.
Article em En | MEDLINE | ID: mdl-38342061
ABSTRACT

PURPOSE:

To determine the diagnostic accuracy of volumetric interpolated breath-hold examination sequences with fat suppression in Dixon technique (VIBE-Dixon) for cardiac thrombus detection.

METHOD:

From our clinical database, we retrospectively identified consecutive patients between 2014 and 2022 who had definite diagnosis or exclusion of cardiac thrombus confirmed by an independent adjudication committee, serving as the reference standard. All patients received 2D-Cine plus 2D-Late-Gadolinium-Enhancement (Cine + LGE) and VIBE-Dixon sequences. Two blinded readers assessed all images for the presence of cardiac thrombus. The diagnostic accuracy of Cine + LGE and VIBE-Dixon was determined and compared.

RESULTS:

Among 141 MRI studies (116 male, mean age 61 years) mean image examination time was 28.8 ± 3.1 s for VIBE-Dixon and 23.3 ± 2.5 min for Cine + LGE. Cardiac thrombus was present in 49 patients (prevalence 35 %). For both readers sensitivity for thrombus detection was significantly higher in VIBE-Dixon compared with Cine + LGE (Reader 1 96 % vs.73 %, Reader 2 96 % vs. 78 %, p < 0.01 for both readers), whereas specificity did not differ significantly (Reader 1 96 % vs. 98 %, Reader 2 92 % vs. 93 %, p > 0.1). Overall diagnostic accuracy of VIBE-Dixon was higher than for Cine + LGE (95 % vs. 89 %, p = 0.02) and was non-inferior to the reference standard (Delta ≤ 5 % with probability > 95 %).

CONCLUSIONS:

Biplanar VIBE-Dixon sequences, acquired within a few seconds, provided a very high diagnostic accuracy for cardiac thrombus detection. They could be used as stand-alone sequences to rapidly screen for cardiac thrombus in patients not amenable to lengthy acquisition times.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Trombose / Meios de Contraste Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Trombose / Meios de Contraste Idioma: En Ano de publicação: 2024 Tipo de documento: Article