Your browser doesn't support javascript.
loading
Three-Dimensional-Printed Instrument for Isothermal Nucleic Acid Amplification with Real-Time Colorimetric Imaging.
Layne, Tiffany R; Scott, Anchi; Cunha, Larissa L; Turiello, Rachelle; Landers, James P.
Afiliação
  • Layne TR; Department of Chemistry, University of Virginia, Charlottesville, VA 22904, USA.
  • Scott A; Department of Chemistry, University of Virginia, Charlottesville, VA 22904, USA.
  • Cunha LL; Department of Chemistry, University of Virginia, Charlottesville, VA 22904, USA.
  • Turiello R; Department of Chemistry, University of Virginia, Charlottesville, VA 22904, USA.
  • Landers JP; Department of Chemistry, University of Virginia, Charlottesville, VA 22904, USA.
Micromachines (Basel) ; 15(2)2024 Feb 14.
Article em En | MEDLINE | ID: mdl-38398999
ABSTRACT
Isothermal amplification methods have become popular in research due to the simplicity of the technology needed to run the reactions. Specifically, loop-mediated isothermal amplification (LAMP) has been widely used for various applications since first reported in 2000. LAMP reactions are commonly monitored with the use of colorimetry. Although color changes associated with positive amplification are apparent to the naked eye, this detection method is subjective due to inherent differences in visual perception from person to person. The objectivity of the colorimetric detection method may be improved by programmed image capture over time with simultaneous heating. As such, the development of a novel, one-step, automated, and integrated analysis system capable of performing these tasks in parallel is detailed herein. The device is adaptable to multiple colorimetric dyes, cost-effective, 3D-printed for single-temperature convective heating, and features an easy-to-use LabVIEW software program developed for automated image analysis. The device was optimized and subsequently validated using four messenger-RNA targets and mock forensic samples. The performance of our device was determined to be comparable to that of a conventional thermal cycler and smartphone image analysis, respectively. Moreover, the outlined system is capable of objective colorimetric analysis, with exceptional throughput of up to 96 samples at once.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article