Your browser doesn't support javascript.
loading
Automated test apparatus for bench-testing the magnetic field homogeneity of NMR transceiver coils.
Uribe, Jose L; Jimenez, Matthew D; Kelz, Jessica I; Liang, Jeanie; Martin, Rachel W.
Afiliação
  • Uribe JL; Department of Chemistry, University of California, Irvine 92697-2025.
  • Jimenez MD; Department of Chemistry, University of California, Irvine 92697-2025.
  • Kelz JI; Department of Chemistry, University of California, Irvine 92697-2025.
  • Liang J; Department of Chemistry, University of California, Irvine 92697-2025.
  • Martin RW; Department of Chemistry, University of California, Irvine 92697-2025.
J Magn Reson Open ; 182024 Mar.
Article em En | MEDLINE | ID: mdl-38444623
ABSTRACT
We describe an automated hands-off bench testing method for measuring the magnetic field profile of transceiver coils for nuclear magnetic resonance (NMR). The scattering parameter (S-parameter) data is measured using a portable network analyzer, and the results are automatically exported to a computer for plotting and viewing. This assay dramatically reduces the time needed to measure the magnetic field (B1) homogeneity profile of a transceiver coil while also improving accuracy relative to manual operation. Here, we demonstrate the method on a saddle coil of a solution-state NMR probe in comparison to profiles obtained using NMR spectroscopy measurements. We also measure the axial and radial homogeneity of a variable-pitch solenoid.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article