Remote monitoring of amyotrophic lateral sclerosis using wearable sensors detects differences in disease progression and survival: a prospective cohort study.
van Unnik, Jordi W J; Meyjes, Myrte; Janse van Mantgem, Mark R; van den Berg, Leonard H; van Eijk, Ruben P A.
Afiliação
  • van Unnik JWJ; Department of Neurology, UMC Utrecht Brain Centre, University Medical Centre Utrecht, Utrecht, the Netherlands.
  • Meyjes M; Department of Neurology, UMC Utrecht Brain Centre, University Medical Centre Utrecht, Utrecht, the Netherlands.
  • Janse van Mantgem MR; Department of Neurology, UMC Utrecht Brain Centre, University Medical Centre Utrecht, Utrecht, the Netherlands.
  • van den Berg LH; Department of Neurology, UMC Utrecht Brain Centre, University Medical Centre Utrecht, Utrecht, the Netherlands.
  • van Eijk RPA; Department of Neurology, UMC Utrecht Brain Centre, University Medical Centre Utrecht, Utrecht, the Netherlands; Biostatistics & Research Support, Julius Centre for Health Sciences and Primary Care, University Medical Centre Utrecht, Utrecht, the Netherlands. Electronic address: r.p.a.vaneijk-2@u
EBioMedicine ; 103: 105104, 2024 May.
Article em En | MEDLINE | ID: mdl-38582030
ABSTRACT

BACKGROUND:

There is an urgent need for objective and sensitive measures to quantify clinical disease progression and gauge the response to treatment in clinical trials for amyotrophic lateral sclerosis (ALS). Here, we evaluate the ability of an accelerometer-derived outcome to detect differential clinical disease progression and assess its longitudinal associations with overall survival in patients with ALS.

METHODS:

Patients with ALS wore an accelerometer on the hip for 3-7 days, every 2-3 months during a multi-year observation period. An accelerometer-derived outcome, the Vertical Movement Index (VMI), was calculated, together with predicted disease progression rates, and jointly analysed with overall survival. The clinical utility of VMI was evaluated using comparisons to patient-reported functionality, while the impact of various monitoring schemes on empirical power was explored through simulations.

FINDINGS:

In total, 97 patients (70.1% male) wore the accelerometer for 1995 days, for a total of 27,701 h. The VMI was highly discriminatory for predicted disease progression rates, revealing faster rates of decline in patients with a worse predicted prognosis compared to those with a better predicted prognosis (p < 0.0001). The VMI was strongly associated with the hazard for death (HR 0.20, 95% CI 0.09-0.44, p < 0.0001), where a decrease of 0.19-0.41 unit was associated with reduced ambulatory status. Recommendations for future studies using accelerometery are provided.

INTERPRETATION:

The results serve as motivation to incorporate accelerometer-derived outcomes in clinical trials, which is essential for further validation of these markers to meaningful endpoints.

FUNDING:

Stichting ALS Nederland (TRICALS-Reactive-II).
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Progressão da Doença / Dispositivos Eletrônicos Vestíveis / Esclerose Lateral Amiotrófica Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Progressão da Doença / Dispositivos Eletrônicos Vestíveis / Esclerose Lateral Amiotrófica Idioma: En Ano de publicação: 2024 Tipo de documento: Article