Your browser doesn't support javascript.
loading
Development of a Nanoplate-Based Digital PCR Test Method for Quantitative Detection of Human Adenovirus DNA.
Dogantürk, Yagmur Eylül; Dag-Güzel, Aylin; Kuskucu, Mert Ahmet.
Afiliação
  • Dogantürk YE; Department of Medical Microbiology, Istanbul University-Cerrahpasa School of Medicine, Istanbul, Türkiye.
  • Dag-Güzel A; Department of Medical Microbiology, Istanbul Aydin University School of Medicine, Istanbul, Türkiye.
  • Kuskucu MA; Department of Medical Microbiology, Istanbul University-Cerrahpasa School of Medicine, Istanbul, Türkiye.
Infect Dis Clin Microbiol ; 5(4): 353-366, 2023 Dec.
Article em En | MEDLINE | ID: mdl-38633848
ABSTRACT

Objective:

Digital polymerase chain reaction (dPCR) assay is an advanced PCR technique that allows for the simultaneous detection and absolute quantification of diverse pathogens.Commercially validated kits available for detecting all subtypes of human adenovirus (HAdV) are limited. This study aimed to demonstrate the development of an in-house nanoplate-based dPCR assay with high sensitivity, even at low copy numbers. Materials and

Methods:

In this methodological study, the standardized HAdV DNA was prepared by amplifying the specific hexon gene region with real-time PCR and purifying the HAdV DNA using magnetic beads from HAdV-positive extractions. Dilutions were tested in triplicate during three independent runs to determine the dynamic range, the limit of detection (LoD), the limit of quantification (LoQ), precision, and reproducibility. The primer and probe sequences used in the study were selected based on a literature review to ensure the detection of all HAdV serotypes in a single run. The selected primers were verified using the US National Center for Biotechnology Information (NBCI) nBLAST tools, and the target sequence was determined using the BioEdit software. The DNA concentration of the stock solution was measured using a Qubit fluorometer. The estimated copy number of the stock solution per milliliter was calculated based on the length of the amplified base sequence and fluorometer measurement.

Results:

The dynamic range of the test was determined to be from 770.4 to 0.9476 cp/µl, with the LoD and LoQ values both being 0.9476 cp/µl. The coefficient of determination (r 2) value of the test was 0.9986.

Conclusion:

The results demonstrated that the dPCR method could be an ideal tool for the diagnosis and absolute quantification of human adenoviruses, especially in low copy numbers. In order to determine the reproducibility of the test and validate the method for field use, it needs to be developed and adapted in various laboratories and supported by clinical studies.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article