Finite element analysis of the effect of framework material and thickness on the biomechanical performance of 'All­on­Four' full-arch prosthesis.
Shash, Yomna H; El-Wakad, Mohamed T; Eldosoky, Mohamed A A; Dohiem, Mohamed M.
Afiliação
  • Shash YH; Department of Biomedical Engineering, Helwan University, Cairo, Egypt.
  • El-Wakad MT; Faculty of Engineering &Technology, Future University in Egypt, Cairo, Egypt.
  • Eldosoky MAA; Department of Biomedical Engineering, Helwan University, Cairo, Egypt.
  • Dohiem MM; Department of Prosthodontics, Zagazig University, Zagazig, Egypt.
Article em En | MEDLINE | ID: mdl-38756021
ABSTRACT
The aim of this research was to evaluate the stress distribution in the 'All-on-Four' prosthesis and the surrounding bone, with different framework materials and thicknesses. Five frameworks (alumina, zirconia, titanium, fiberglass reinforced resin (FRR), and polyether ether ketone (PEEK)) with two thicknesses (3.5 &5.5 mm) were stimulated in this research. A vertical force of 200 N was applied on a 1 mm circular area, at the cantilever, and at the region of the incisors, simulating different mastication mechanisms. The results illustrated that the 5.5 mm framework reduced the stresses on most parts, mucosa, and bone tissues, compared to 3.5 mm.